SENSE技术在下肢静脉成像中的应用

Evaluation of SENSE MRV technique in lower limb veins
下载PDF
导出
摘要 目的:探讨磁共振SENSE技术在下肢静脉成像中的应用价值。方法:不用造影剂,采用非SENSE线圈及技术,SENSE线圈及技术对34例临床怀疑下肢静脉病变的患者同时进行磁共振静脉成像并分析其应用价值。结果:所有病例均较清晰显示下肢静脉,应用SENSE技术的下肢静脉图像与非SENSE技术的图像质量相同,扫描时间减少一半。在扫描范围及时间不变时提高分辨率的参数,可明显提高图像质量。结论:SENSE技术是一种可以大大减少扫描时间,不影响空间分辨率的新方法,在磁共振快速成像及临床中的应用前景广阔。 Objective: To explore the application value of the technique of SENSE MRV in lower limb veins. Methods: Without SENSE technique and with SENSE technique were performed in 34 patients suspected with venous diseases of the lower limb. Contrast enhancement were not used in all examinations. Results: All cases acquired excellent image and anatomic changes in the lower limb veins. The scan time with SENSE technique of reduction factor 2.0 in lower limb veins imaging decreased about 50% of the counterparts without SENSE technique and the contrast resolution on SENSE imaging almost kept the same as that on conventional sequence. Changing the parameters which enhance resolution as the same as scanning scope and time may obviously improve the picture quality. Conclusion: It offers a new approach to reduce the scan time at preserved spectral resolution, SENSE technique shows a widen promising in fast MRI and clinical application.
出处 《中国临床医学影像杂志》 CAS 北大核心 2007年第5期322-323,326,共3页 Journal of China Clinic Medical Imaging
关键词 静脉 下肢 磁共振成像 veins lower extremity magnetic resonance imaging
  • 相关文献

参考文献7

二级参考文献18

  • 1[1]Pennell D. Cardiovascular magnetic resonance[J].Heart.,2001,85(5):581-589.
  • 2[2]Steenbeck J, Pruessmann K. Technical developments in cardiac MRI: 2000 update[J].Rays,2001,26(1):15-34.
  • 3[3]Pruessmann KP, Weiger M, Boesiger P. Sensitivity encoded cardiac MRI[J].J Cardiovasc Magn Reson,2001,3(1):1-9.
  • 4[4]Pruessmann KP, Weiger M, Scheidegger MB, et al. SENSE: sensitivity encoding for fast MRI[J].Magn Reson Med. 1999 ,42(5):952-962.
  • 5[5]Bammer R, Keeling SL, Augustin M, et al. Improved diffusion-weighted single-shot echo-planar imaging (EPI) in stroke using sensitivity encoding (SENSE)[J].Magn Reson Med,2001,46(3):548-554.
  • 6[6]Lima JAC, Judd RM, Bazille A. Regional heterogeneity of human myocardial infarcts demonstrated by contrast-enhanced MRI: potential mechanisms[J].Circulation,1995,92(5):1117-1121.
  • 7[7]Weiger M,Pruessmann KP,Boesiger P. Cardiac real-time imaging using SENSE. Sensitivity encoding scheme[J].Magn Reson Med,2000,43(2):177-184.
  • 8[1]Dixon W T. Simple proton spectroscopic imaging. Radiolog, 1984,153: 189- 194
  • 9[2]Mansfield P. Spatial mapping of the chemical shift in NMR. Magn Reson Med, 1984, 1:370-386
  • 10[3]Adalsteinsson E, Irarrazabal P, Topp S, et al. Volumetric spectroscopic imaging with spiral-based k-space trajectories. Magn Reson Med, 1998,39: 889 - 898

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部