摘要 根据最新国际电工委员会标准IEC62220-1,我们对我院数字X线平板探测器系统的物理特性进行测试。设计建立符合IEC标准的实验测试设备和测试方法,并对某些测试条件对测量结果的影响进行初步研究。 DQEs for a digital X-ray flat panel detector system has been measured in accordance with the new international standard IEC 62220 - 1. A measurement set - up complying with IEC 62220 - 1 has been realized, the influence of measuring set- up was evaluated in this paper
出处 《医疗设备信息》 2007年第4期1-4,共4页 Information of Medical Equipment
基金 国家自然科学基金资助(30470531)
关键词 数字X射线机 DQE调制传递函数 噪声功率谱 噪声等价量子数 digiter X-ray machine modulation transfer function noise power spectrum noise equivalent number of quanta
  • 相关文献

参考文献10

  • 1Cunningham IA.Applied linear-systems theory,in:Handbook of medical imaging,vol.1:Physics and psychophysics[J].Beutel,H.L.Kundel,R.L.Van Metter,Editors,Bellingham:SPIE Press,2000:79.
  • 2Samei E,Flynn MJ.An experimental comparison of detector performance for computed radiography systems[J].Medical Physics,2002,29:(4),447-459.
  • 3Samei E,Flynn MJ.An experimental comparison of detector performance for direct and indirect digital radiography systems[J].Medical Physics,2003,30:608-622.
  • 4Bath M.Sund P,Mansson LG.Evaluation of the imaging properties of two generations of a CCDbased system for digital chest radiography[J].Medical Physics 2002,29(10),2286-2297.
  • 5IEC 62220-1:Medical electrical equipmentCharacteristics of digital X-ray imaging devices:Part 1:Determination of the detective quantum efficiency[J].International Electrotechnical Commission (IEC),Geneva,Switzerland,2003.
  • 6Illers H,Buhr E,Bergmann D,C.Hoeschen."Measurement of the detective quantum efficiency (DQE) of digital x-ray imaging devices according to the forthcoming standard IEC 62220-1"[J].Proceeding of SPIE Medical Imaging,2004.
  • 7Buhr E,Gunther-Kohfahl S and Neitzel U.Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image[J].Medical Physics,2003,30(9):2323-2331.
  • 8黄岗,严汉民.点扩散函数测量CT机MTF的应用研究[J].医疗设备信息,2004,19(8):10-12. 被引量:5
  • 9白玫,彭明辰.X线CT成像质量评估中的调制传递函数的测试与分析[J].医疗设备信息,2006,21(5):11-13. 被引量:15
  • 10刘景鑫,周安国,刘更,刘建国,孟凡君.CR成像空间分辨率的实验研究[J].医疗设备信息,2005,20(7):10-11. 被引量:3

二级参考文献13

  • 1邵东宁.CR系统在床边胸部X线摄影中的应用与特性研究[J].医疗设备信息,2004,19(10):41-42. 被引量:1
  • 2胡曙光,昌仁民,冯小刚,康立丽,林意群,张雪林.CT的MTF测量和评价[J].中国医学物理学杂志,1997,14(1):30-32. 被引量:7
  • 3Specification and acceptance testing of computed tomography scanners[R].AAPM report No.39,1993.
  • 4Boone JM. Determination of the presampled MTF in computed tomography[J].Medical Physics,2001,28 (3):356-360.
  • 5Goodenough D,Atkins F,Dyer S.Automated techniques for quality assurance of radiological image modalities.SPIE.Image Capture,Formatting,and Display,1991,1444:87-99.
  • 6Dainty JC,Shaw R.Image Science[M].New York:Academic Press,1974.
  • 7Goodenough D,Atkins F.Theoretical and practical aspects of automated quality assurance approaches,particularly for CT[C].Proceedings of the ICR 89,editors Silve,et.Al,1990.
  • 8Fujita H,Tsai D-Y,Itoh T,et al.A simple method for determining the modulation transfer function in digital radiography[J].IEEE Transactions on Medical Imaging,1992,11:34-39.
  • 9Cunningham IA,Reid BK.Signal and noise in modulation transfer function determinations using the slit,wire,and edge techniques[J].Medical Physics,1992,19:1037-1044.
  • 10Samei E,Flynn MJ,Reimann DA.A method for measurement the presampled MTF of digital radiographic systems using an edge test device[J].Medical Physics,1998,25:102-113.