不同角膜厚度下动态轮廓眼压计和Goldmann压平眼压计在不同角膜厚度的测量值比较 被引量:10

Comparison of dynamic contour tonometry with Goldmann applanation tonometry in measurement of different central corneal thickness
下载PDF
导出
摘要 目的比较动态轮廓眼压计(dynamic contour to-nometry,DCT)和Goldmann压平眼压计(Goldmann applanation tonometry,GAT)在中央角膜厚度(central corneal thickness,CCT)不同的正常眼测量值的差异。方法患者69例(69眼)按CCT不同分为正常组(520~580μm)42眼、变薄组(450~520μm)16眼、增厚组(580~640μm)11眼。各组分别用DCT和GAT测量眼压。结果正常组平均CCT为(574·55±16·40)μm,平均GAT值为(16.62±3.15)mmHg(1kPa=7.5mmHg),平均DCT值为(16.78±2.86)mmHg,DCT与GAP差异无统计学意义(P>0.05),平均GAT-DCT值为(-0.16±0.96)mmHg;变薄组平均CCT为(499.44±16.70)μm,平均GAT值为(14.03±2.72)mmHg,平均DCT值为(17.06±2.72)mmHg,2者差异有统计学意义(P<0·01),平均GAT-DCT值为(-3.03±1.32)mmHg;增厚组平均CCT为(605·18±17·90)μm,平均GAT值为(16.91±3·94)mmHg,平均DCT值为(16·80±4.25)mmHg,2者差异无统计学意义(P>0.05),平均GAT-DCT值为(0.05±2·47)mmHg。结论在角膜厚度正常眼,DCT值和GAT值较一致;在角膜厚度变薄眼,前者较后者准确;在角膜厚度变厚眼未体现DCT的优越性。 Objective To compare the value between dynamic contour tonometry(DCT) and Goldmann applanation tonometry (GAT) in measurement of different central corneal thickness (CCT). Methods According to CCT, 69 patients ( 69 eyes ) were divided into three groups: normal group (from 520 μm to 580 μm,42 eyes), thin group( from 450 μm to 520 μm, 16 eyes), and thick group ( from 580 μm to 640 μm, 11 eyes ). Intraocular pressure was measured in each group with GAT and DCF. Results In normal group, in which the mean CCT was ( 574.55 ± 16.40 )μm , the mean value of GAT was ( 16.62 ±3.15) mmHg( 1 kPa = 7.5 mmHg) and DCT was (16.78 ±2.86)mmHg,there was no statistical difference( P 〉 0.05 ). In thin group, in which the mean CCT was ( 499.44 ±16.70 )μm, the mean GAT was ( 14.03 ± 2.72 ) mmHg and DCT was ( 17.06 ± 2.72 ) mmHg, there was statistical difference ( P 〈 0.01 ). In thick group, in which the mean CCT vas (605.18± 17.90) μm,the mean value of GAT was (16.91 ± 3, 94) mmHg and ± was( 16.80 ± 4.25 ) mmHg, there was no statistical difference( P〉0.05 ). The mean GAT - DCT difference was ( - 3.03 ± 1.32 ) mmHg in thin group and (0.05 ± 2.47 ) mmHg in thick group. Below 520 μm,a reduction of 10 μm in CCT resulted in a significant underestimation of the value of GAT by 0. 67 mmHg (P 〈 0. 05 ). Conclusion In normal eyes, the value of DCT is equal to that of GAT. In eyes with thin cornea, the value of DCT may give a more accurate assessment than that of GAT, but it does not appear to have any benefit over GAT in eyes with thick cornea.
出处 《眼科新进展》 CAS 2007年第5期363-365,共3页 Recent Advances in Ophthalmology
关键词 中央角膜厚度 动态轮廓眼压计 Glodmann压平眼压计 central corneal thickness dynamic contour tonometry Goldmann applanation tonometry
  • 相关文献

参考文献10

  • 1Ehlers N,Bramsen T,Sperling S.Applanation tonometry and central corneal thickness[J].Acta Ophthalmol (Copenh) 1975 ;53:34-43.
  • 2詹素华,庞国祥,金玉梅,孙玉敏.角膜中央厚度与眼压读数[J].眼科新进展,2001,21(2):126-127. 被引量:12
  • 3Duch S,Serra A,Castanera J.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma 2001; 10:261-265.
  • 4Siganos DS,Papasterigiou GI,Moedas C.Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK[J].J Cataract Refract Surg 2004;30:746-751.
  • 5Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci 2003 ;44:3790-3794.
  • 6Zadok D,Raifkup F,Landao D.Intraocular pressure after LASIK for clinical research[J].Ophthalmology 1996; 103:183-185.
  • 7张雪翎,王丹梅,孟觉天,金红颖.LASIK后眼压测量值变化分析[J].眼科新进展,2003,23(6):401-401. 被引量:4
  • 8Kniestedt C,Nee M,Stamper RL.Dynamic contour tonometry:a comparative study on himan cadaver eyes[J].Arch Ophthalmol 2004; 122:1287-1293.
  • 9Kniestedt C,Nee M,Stamper RL.Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states[J].Graefe Arch Clin Exp Ophthalmol 2004 ;243:359-366.
  • 10Doyle A,Ophth FRC,Lachkar Y.Comparison of dynamic contour tonometry with Goldmann applanation tonometry over a wide range of central corneal thickness[J].J Glaucoma 2005; 14:288-292.

二级参考文献3

  • 1Schipper I ,Senn P, Niesen U. Are we measuring the right intraocular pressure after excimer laser photorefractive laser keratoplasty in myopia[J]?Klin Monatsbl Augenheilkd 1995;206:322-324.
  • 2Chatterjee A,Shah S,Bessant DA,Naroo SA,Doyle SJ.Reduction in intraocular pressure after laser photorefractive keratectomy.Correlation with pretreatment myopia [J].Ophthalmology 1997;104:355-359.
  • 3潘怡,张泳,廉井财,王康孙.准分子激光原位角膜磨镶术后眼压和角膜厚度的研究[J].中华眼科杂志,1999,35(5):359-362. 被引量:31

共引文献14

同被引文献128

引证文献10

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部