两种探测锉电测根管长度稳定性和准确性的比较 被引量:1

COMPARISON OF THE STABILITY AND THE ACCURACY OF ELECTRONIC ROOT CANAL LENGTH MEASUREMENT USING TWO KINDS OF DETECTING FILES
下载PDF
导出
摘要 目的比较不锈钢根管锉和ProTaper镍钛手用根管锉电测根管长度的稳定性和准确性。方法选择临床需行根管治疗、根尖发育完全的60颗活髓后牙为研究对象。开髓、拔髓后探通根管,在同一颗牙上先用不锈钢锉初尖锉作为NeosonoTM根管长度电测仪的探测电极测量根管长度,评价其稳定性,测量稳定的根管则带针拍X光数字成像(radio visio graphy,RVG)牙片评价其准确性,然后用ProTaper手用镍钛初尖锉测量根管长度,同样的方法评价其稳定性和准确性,将结果进行统计分析。结果①不锈钢组EAL(根管长度电测仪)测量稳定的根管数为102个,占65.38%,镍钛组为142个,占34.62%,两者相比有显著性差异。②在两种探测锉测量均稳定的根管中,不锈钢组EAL测量准确的根管数为89个,占87.25%,镍钛组为95个,占93.14%,两者相比无显著性差异。结论大锥度镍钛探测电极下EAL测量稳定性高;探测电极的锥度、材质对EAL测量准确性影响很小。临床上使用大锥度镍钛锉电测根管长度更稳定可靠,且安全准确。 Objective To examine the stability and the accuracy of electronic root canal length measurement (ERCLM) using stainless steel files and ProTaper NiTi files clinically. Methods 60 posterior teeth which need endodontic treatment were studied clinically, after gaining access to their roots, The canal length (IL) was measured with the NeosonoTM apex locator by negotiating a size 10 stainless steel file or ProTaper NiTi files ( S1 or S2 ) to the apical constriction,respectively . The stability and accuracy of the results of electronic root canal length measurement were evaluated and verified radiographically (RVG) in two groups. Result The ERCLM were more stable in Pro- Taper nickel -titanium group 142 (34.62%)than in stainless steel group 102 (65.38%) (p 〈 0.05 ), while the num- bers of precise Electronically measured lengths (EML) was 89(87.25% ) in stainless steel group and 95(93.14% ) in nickel - titanium group. There was no statistically significant between two groups (p 〉 0.05 ) . Coulusion ER- CLM is more stable by using big taper nickel - titaniume exploring electrode than using stainless steel files, while the effect on the accuracy can not influenced by the taper and the material of exploring electrode.
作者 张杨 孙书昱
机构地区 广东省口腔医院
出处 《现代医院》 2007年第6期38-39,共2页 Modern Hospitals
关键词 探测锉 电测根管长度 稳定性 准确性 detecting file, electronic root canal length measurement( ERCLM), stability, accuracy
  • 相关文献

参考文献6

  • 1KAUFMAN A Y,KEILA S,YOSHPE M.Accuracy of a new apex locator[J].an in vltm study,2002,35(2):186-192.
  • 2ELAYOUTI A,WEIGER R,LOST C.The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working[J].JEndod,2002,28(2):116-119.
  • 3PRATIEN D H,MCDONALD N J.Comparison of radiographic and electronic working lengths[J].J Ended,1996,22(4):173-176.
  • 4唐晓琳,潘春玲,王贤辐.两种牙齿操作长度电测仪准确性的体外研究[J].牙体牙髓牙周病学杂志,2002,12(10):547-549. 被引量:12
  • 5ANTHONYS,GARY R,PETER C.The accuracy of the root ZX eleetmnic apex locator using stainless-steel and nickel-titanium files[J].J Ended,2003,29(10):662-663.
  • 6HUANG L.An experimental study of the principle of electronic root canal measurement[J].J Endodon,1987,13(2):60-64.

二级参考文献8

  • 1黄力子.根管长度电测法及其原理[J].中华口腔医学杂志,1988,23(5):297-297.
  • 2Kovacevic M, Tamarut T. Influence of the concentration of ions and foramen diameter on the accuracy of electronic root canal length measurement-an experimental study[J]. J Endod, 1998,24(5):346
  • 3Weiger R, John C, Geigle H,et al. An in vitro comparison of two modern apex locators[J]. J Endod,1999,25(11):765
  • 4Hung L . An experimental study of the principle of electronic root canal measurement[J]. J Endod,1987,13(2):60
  • 5Russell J, Michael S,Jerome C . An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices[J]. J Endod,1994,20(12):605
  • 6Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator[J]. J Endod, 1998,24(6):438
  • 7Coleman CL,Svec TA. Analysis of Ni-Ti versus stainless steel instrumentation in resin simulated canals[J]. J Endod, 1997,23(4):232
  • 8付少平,乐进秋,范兵.根管长度电测法的临床应用[J].国外医学(口腔医学分册),2001,28(3):141-143. 被引量:16

共引文献11

同被引文献5

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部