小动物正电子断层成像仪性能测试 被引量:2

Performance evaluation of the microPET system
原文传递
导出
摘要 设计了一系列实验对Concorde Micro PETRodent R4小动物正电子断层成像系统(该系统是Concorde Microsystems公司开发的小动物专用PET扫描仪,该扫描仪具有32层探测器环,动物入口孔径为120mm,横向视野为100mm,轴向深度为78mm,标称的视野中央处的空间分辨率可达1.8mm以下)的空间分辨率、灵敏度、散射率等性能参数进行了测试,测试结果显示该扫描仪视野中心的横向空间分辨率为1.9mm,轴向空间分辨率为1.88mm,在能窗设定为250-750keV时系统的绝对灵敏度为39.88cps/kBq,同样能窗下二维重建散射分数为50.6%,三维重建散射分数为32.3%. The Concorde MicroPET Rodent R4 system designed by Concorde Microsystems Inc. is the dedicated positron emission tomography (ECT) for studies of rodents. The scanner has 32 detector rings and a 120mm animal aperture. The transversal field of view(FOV) is 100mm, and the axial FOV is 78 mm. The spatial resolution at the center of FOV is smaller than 1.8 mm. This paper designs a series of experiments to evaluate the performance parameters (spatial resolution, sensitivity, scatter ets. ) of the scanner. In the center FOV the transversal spatial resolution is 1 .gmm, the axial spatial resolution is 1.88mm. The absolute sensitivity of the system is 39.88 cps/kBq for an energy window of 250KeV-750KeV. For the same energy window the 2D-reconstructed scatter fractions is 50.6%, the 3D-reconstructed scatter fractions is 32.3 %.
出处 《物理学报》 SCIE EI CAS CSCD 北大核心 2007年第7期4229-4234,共6页 Acta Physica Sinica
基金 国家重点基础研究发展计划(批准号:2003CB716100) 浙江省科技计划重点项目(批准号:2005C21019) 浙江大学现代光学仪器国家重点实验室开放课题(批准号:LMOI-0403)资助的课题.~~
关键词 正电子断层成像仪 性能测试 PET扫描仪 microPET, performance evaluation
  • 相关文献

参考文献10

  • 1Coneorde Mierosystems Inc.Coneorde Microsystems manual 2006
  • 2Knoess C,Siegel S,Smith A,Newport D,Richerzhagen N,Winkeler A,Jacobs A,Goble R N,Graf R,Wienhard K,Heis W D 2003 Eur J Nucl Med Mol Imaging 30 737
  • 3田毅.2006.浙江大学博士论文
  • 4GB/T18988.1-2003[正电子发射断层成像装置国家标准]
  • 5IEC 61675-1:1998[正电子发射断层成像装置国家标准]
  • 6Tai Y C,Chatziioannou A,Siegel S,Young J,Newport D,Goble R N,Nutt R E,Cherry S R 2001 Phys.Med.Biol 46 1845
  • 7Tai Y C,Ruangma A,Rowland D,Siegel S,Newport D,Chow P L,Laforest R 2005 the Journal of Nuclear Medicine 46 455
  • 8NEMA NU 2-2001 Standard
  • 9赵辉,黄健,蓝海,董宝中.大鼠骨质疏松模型的小角x射线散射研究[J].物理学报,2004,53(6):2005-2008. 被引量:1
  • 10魏强,刘海,何世禹,郝小鹏,魏龙.质子辐照铝膜反射镜的慢正电子湮没研究[J].物理学报,2006,55(10):5525-5530. 被引量:5

二级参考文献21

  • 1胡家升,张景和,王宝林,孙可煦,马洪良,徐和平.X光掠入射平面反射镜系统的制备[J].中国激光,1995,22(4):276-280. 被引量:2
  • 2Landis W J, Paine M C and Glimcher M J 1977 J. Ultrastruct Res.59 217
  • 3Nelson D G A and Barry J C 1989 Anat Rec. 224 265
  • 4Landis W J et al 1990 J. Cell. Biol. 111 24a
  • 5Landis W J et al 1977 J. Ultrastruct Res. 59 185
  • 6Ascenzi A et al 1985 Calcif. Tissue Int. 37 659
  • 7Arsenault A L and Grynpas M D 1988 Calcif. Tissue Int. 43 219
  • 8Sasaki N et al 1989 J. Biomech 22 157
  • 9Fratzl P et al 1992 Bone Miner Res. 7 329
  • 10Fratzl P et al 1991 Calcif . Tissue Int. 48 407

共引文献4

同被引文献15

  • 1何义杰,陈英茂,邵明哲,张书文,尹大一,徐白萱,王瑞民,姚树林,丁勇,关志伟,田嘉禾.能窗下限对三维PET图像质量的影响[J].中华核医学杂志,2004,24(5):305-307. 被引量:3
  • 2周伟,尹端沚,汪勇先.小动物PET[J].核技术,2006,29(3):207-213. 被引量:11
  • 3Adam L E, Karp J S, Brlx G 1999 Phys. Med. Biol. 44 2879
  • 4Jan S, Santin G, Strul D 2004 Phys. Med. Biol. 49 4543
  • 5Assie K, Breton V, Buvat I, Comtat C 2004 NIMA 527 180
  • 6Watson C C 2000 IEEE Trans. Nucl. Sci. NS-47 1587
  • 7Gao F, Yamada B, Watanabe M 2008 Japanese Journal of Medical Physics 28 ( sup 2) 2 49
  • 8Lubberink M, Kosugi T, Schneider H, Ohba H, Bergstrom M 2004 Phys. Med. Biol. 49 833
  • 9Markiewicz P J, Tamal M, Julyan P J, Hastings D L, Reader A J 2007 Phys. Med. Biol. 52 829
  • 10Gopal B. Basics of PET Imaging[M]. New York: Springer, 2010.

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部