摘要 目的研制、评价血清铁Nitro-PAPS自动分析法试剂盒。方法在表面活性剂存在下,用2-(5-硝基-2-吡啶偶氮)-5-[N-正丙基-N-(3-磺酸丙基)氨基]苯酚二钠(Nitro-PAPS)作显色剂双试剂两点终点法测定血清铁。结果该方法显色络合物最大吸收波长为591 nm,线性范围达85μmol/L,表观摩尔吸光数为1.11×105L.mol-1.cm-1,回收率为98.5%-104.1%,批内变异系数(CV)和批间变异系数分别为0.021和0.035,与原子吸收分光光度法(AAS)比较相关良好,Y=0.997X-0.223,r=0.9786,P>0.05,224例健康人血清铁含量为(10.83-29.07)μmol/L(±2 s)。结论用本试剂盒双试剂自动化分析法测定血清铁方法简便、灵敏可靠,适合临床应用。 Objective To investigate and evaluate the the reagent kit of automatic analysis the zinc in serum.Methods For determination of serum zinc with Nitro-PAPS in the presence of surfactant. Results The method was liner in the range of 0 - 85μmol/L, the rate of recovery was 98.5 % - 104.1%,, the withinrun and between run CV was 0. 021 and 0. 035. Comparing this method (Y) with AAS (X),we obtained the regression equation: Y = 0. 997 X-0. 223, r = 0. 9786, P 〉 0.05. The reference interval(± 2s) was 10.83- 29.07μmol. L^-1 ( n = 224) Conclusion This reagent kit of automatic analysis the iron in serum is simple, sensitive and suit-able for clinical application.
出处 《中国实验诊断学》 2007年第7期928-929,共2页 Chinese Journal of Laboratory Diagnosis
关键词 血清 自动分析法 试剂盒 Nitro-PAPS iron serum automatic analysis reagent kit Nitro-PAPS.
  • 相关文献

参考文献7

  • 1Benzo ZA,Fraile R,Gomez C,et al.Determination of serum iron by aas:deproteinization versus direct analysis[J].Clin Chim Acta,1991,197(2):141.
  • 2Traister GL,Schilt AA.Water-soluble sulfonated chromogenic reagents of the ferroin type and determination of iron and copper in water,blood serum,and beer with the tetraammonium salt of 2,4-bis(5,6-diphenyl-1,2,4-triazin-3-yl)pyridinetetrasulfonic acid[J].Anal Chem,1976,48(8):1216.
  • 3Escobar R,Cano Pavon JM.Selective spectrophotometric determination of trace amounts of iron with di(2-pyridyl)-NN-di[(8-quindyl)amino] methane:determination of into in blood serum[J].Analyst,1983,108(1288):821.
  • 4Charlier C,Plomteux G,Vernet M,et al.Modification of the selected method for the determination of serum iron.Substitution of bathophenanthroline by ferene S[J].Ann Biol Clin,1992,50(3):191.
  • 5连国军,李东,方芳,陈红.血清铁的自动化分析法[J].上海医学检验杂志,2001,16(2):127-127. 被引量:15
  • 6曹建明,肖洪武,姜红鹰,沈财成,方芳,李敏.新的高灵敏分光光度法在血清铁测定中的应用[J].中华检验医学杂志,2004,27(3):137-139. 被引量:8
  • 7陶玉年,郭立新.双缩脲试剂对血清铜测定有携带污染[J].临床检验杂志,2002,20(1):41-41. 被引量:29