期刊文献+

基于振动法和柯氏音法的无创血压测量值差异性研究 被引量:2

Study on difference of noninvasive blood pressure measurement method based on oscillometric and auscultatory
下载PDF
导出
摘要 对无创血压测量仪测量值误差异性展开分析。用与各种血压测量技术进行比较的方法,着重分析了振动法无创血压测量和柯氏音听诊法。使用振动法无创伤血压测量监护仪和柯氏音听诊法对样本人群进行了血压测量,分析所得数据,结果表明,基于振动法和柯氏音法无创血压测量仪测量值差异小,准确性也比较好。同时,也展望了无创血压装置未来的发展方向。 This paper introduces and compares several blood pressure measurements, among which two noninvaslve blood pressure measurements based on oscillometric and auscuhatory methods are mainly compared. It also analyses the characteristics and advantages of oscillometric method as well as introduces a noninvasive blood pressure measurement monitor based on oscillometric. By using this monitor and a common auscultory method, we conducted a public survey. The result for statistical data analysis showed that the difference between oseillometric and auscuhatory methods are quite small, and the accuracy of them are both quite good .This paper also previews the future of the noninvasive blood pressure measurement
出处 《医疗卫生装备》 CAS 2007年第7期29-30,共2页 Chinese Medical Equipment Journal
基金 重庆大学大学生创新基金资助项目 重庆大学生物工程学院创新人才培养基金项目资助
关键词 无创血压测量 柯氏音法 振动法 差异性 noninvasive blood pressure measurements auscultatory method oseillometric method difference
  • 相关文献

参考文献7

二级参考文献43

  • 1邹峰.脉搏波信号处理的一种简单可靠的实时算法[J].生物医学工程学杂志,1994,11(2):182-183. 被引量:2
  • 2扬子彬.当代医学新理论与新技术丛书:基础医学卷[J].生物医学工程学,2000,.
  • 3陈芝村.动脉血压调节机理及测量[M].北京人民军医出版社,1987..
  • 4K.Yanakoshi et al. Indirect Measurement of instantaneous Arterial Blood Pressure in the Human Finger by the Vascular Unloading Technique. IEEE Transa onBi0m Engi.vol.BME-27. Mar 1980, 3: 150-155.
  • 5Ramsey M. Noninvasive automatic determination of mean arterial pressure.Med. Biol. Eng Comput.1979, 17(1): 11-17.
  • 6Mauck, G.W, et al. The meaning of the points of maximum oscillations in cuff pressure in the indirect measurement of blood pressure-Part Ⅱ. J.Biomech. Eng.1980, 102:28-33.
  • 7J.Penaz. Photoelectnc measurement of blood pressure,volume and flow in the finger, in Dig. 10th int Conf.Med.BioI.End..., Dresden, Germany, 1973, 104.
  • 8Imholz BP et al. Fifteen years experience with finger artedal pressure monitoring:assessment of the technology. Cardiovasc Res. 1998, 38(3):605-616.
  • 9Spainski A. Standard algorithm of blood_pressure measurement by oscillometertc method Med. &Biol. &Comput. 1992, (3): 671.
  • 10Spainski A. Standard algorithm of blood_pressure measurement by oscillometadc method. Med, &Biol. &ComDut. 1993: 31: 204.

共引文献62

同被引文献11

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部