摘要 研究了数字图像消卷积在显微测量中的应用。在测量系统点扩展函数的基础上,利用频谱空间变换消卷积图像处理方法,消除了传统光学显微成像中的几何不清晰度。通过对显微图像进行复原,实现了反射式一维线宽的显微精确测量。在医用接骨钢板表面质量的评价中,对10μm量级的划痕宽度实现了精确和方便的测量。该技术对相关计量技术的发展具有重要意义。 Application of deconvolution of digital images in macroscopic measurement is studied. On the basis of measuring the point spread function of the system, a method of transformation of spectrum space is used in order to obtain clearer images, without the geometric blurring in traditional microscopy, and the width d one-dimensional line can be precisely measured. The technique is used to evaluate the surfaces of steel plates applied in osteopathy, and the scratches scaled in 10μm are measured accurately and easily. The technique makes great sense in the development of corresponding measurement.
出处 《光学技术》 EI CAS CSCD 北大核心 2007年第4期576-579,共4页 Optical Technique
关键词 显微测量 数字成像 消卷积 点扩展函数 macroscopic measurement digital imaging deconvolution point spread function
  • 相关文献

参考文献13

  • 1杨国光.近代光学测试技术[M].杭州:浙江大学出版社,2001..
  • 2麦伟膦.光学传递函数及其数理基础[M].北京:国防工业出版社,1979.
  • 3王植,贺赛先,毛庆洲,仲思东.数字图像处理技术在钢坯在线检测系统中的应用[J].武汉大学学报(信息科学版),2005,30(3):269-273. 被引量:7
  • 4Ania Majewska, Glen Yiu, Rafael Yuste. A custom-made two-photon microscope and deconvolution system[J]. Pflugers Arch - Eur J Physiol, 2000, 441 : 398-408.
  • 5中华人民共和国医药行业标准YY 0345-2002,骨接合植入物(金属骨针)[S].
  • 6刘维一,于德月,王肇圻,母国光.用迭代法消除数字图像放大后的模糊[J].光电子.激光,2002,13(4):398-400. 被引量:9
  • 7江月松.直线阵光学综合孔径成像中的子孔径尺寸效应[J].光学学报,2005,25(8):1042-1047. 被引量:14
  • 8Jean-Baptiste Sibarita. Deeonvolution Microscopy[J ]. Adv Bioehem Engin/Biotechnol, 2005,95: 201-243.
  • 9Park J S, Kihm K D. Three-dimensional micro-PTV using deconvolution microscopy[J]. Experiments in Fluids,2005.
  • 10Oozeer N, N Udaya Shankar. Limitations of the tangent plane approximation for wide-field imaging using the mauritius radio telescope[J]. Astrophysics and Space Science, 2002,282:43-55.

二级参考文献43

  • 1刘诚 姜锦虎 等.数字散斑相关测量系统抗噪声干扰能力的判据[J].上海力学,1996,17:110-113.
  • 2肖全标 姜锦虎 等.数字散斑相关测量中噪声的实验研究[J].上海力学,1996,17:114-115.
  • 3Yamaguehi. Simplified laser-speckle strain gauge [J]. Opt.Engng, 1982, 21(3): 436-440.
  • 4W. H. Peters, W. F. Ranson. Digital imaging techniques in experiment[J]. Opt. Engng, 1982, 21(3): 427-431.
  • 5姜锦虎 王海凤 刘诚.数字散斑图跟相关测量系统抗噪声抗干扰能力关系的研究[C]..现代力学测试技术[C].华南理工大学出版社,1998.294-297.
  • 6[1]Castleman K R. Digital Image Processing. Beijing: Tsinghua & Prentice Hall,1996
  • 7[2]McNally J G, Karpova T, Copper J et al.Three dimensional imaging by deconvolution microscopy. Methods, 1999,19:373~385
  • 8[3]Shaw P J. Comparison of wide-field/ deconvolution and confocal microscopy for 3D imaging. In:Pawley J B ed. Handbook of Biological Confocal Microscopy. New York: Plenum Press, 1995
  • 9[4]McNally J G, Markham J, Conchello J A. Comparing methods for 3D microscopy. SPIE, 1998,3216:108~116
  • 10[5]Gibson D F, Lanni F. Experimental test of an analytical model of aberration in an oilimmersion objective lens used in three- dimensional light microscopy. J Opt Soc Am,1991,A8(10): 1601~1613