摘要 采用光子计数法研制大气SO_2浓度分析仪,计数结果对应SO_2的浓度值。光子计数器的计数随测量环境温度的变化有一定波动。为消除这一影响,需要对光子计数器的工作环境温度进行测控,测控系统采用半导体制冷片来降低光子计数器的环境温度,并实时显示计数器的当前温度。实验结果证明,在室内温度下,采用级联制冷法,仪器在开启25 min后,光子计数器温度能达到5℃以下。 The principle of photon counter is applied to the development of the analyzer of SO2 concentration in air. The result read by the analyzer corresponding to the concentration of SO2 fluctuates as the environment temperature varies. Necessarily, the environment temperature of photon counter should he monitored and controlled on purpose to eliminate the negative fluctuation. The system adopts semiconductor refrigerator to reduce the circumstance temperature of photon counter, and display temperature of counter in real-time. The experiment result shows that, under indoor temperature, the temperature of photon counter is lower than 5 ℃ after the analyzer working for 25 minutes when the method of connection refrigerate is adopted.
出处 《国外电子测量技术》 2007年第7期35-37,共3页 Foreign Electronic Measurement Technology
基金 广州市环保工程示范项目资助(穗计园[1999]31号)
关键词 SO2浓度监测 温度测量 光子计数 detection of SO2 concentration temperature detector photon counting
  • 相关文献

参考文献5

  • 1OKABE H, SPLITSTONE P L, JOSEPH J. Ball Ambient and source SO2 detector based on a fluorescence method[J]. Journal of the Air Pollution Control Association, 1973, 23(6): 514-516.
  • 2HAMAMATSU PHOTONICS K K. Electron Tube Center. PHOTON COUNTING Using Photomultiplier Tubes[Z]. Japan. : HAMAMATSU , 1998.
  • 3HAMAMATSU PHOTONICS K K. Electron Tube Center. M7824 Photon Counting Board INSTRUCTION MANUAL[Z]. Japan: HAMAMATSU 1999.
  • 4王戴,吕强中,王同峰.温度传感器在数据采集系统中的应用[J].计算机测量与控制,2004,12(1):95-97. 被引量:8
  • 5张红,黄志辉.AD590在轴承温升测试中的精度分析[J].航空计测技术,2003,23(5):24-26. 被引量:1