摘要
[目的]对比研究正常股骨颈、单纯髓芯减压、髓芯减压+钛合金支撑架植入治疗实验性狗股骨头坏死对股骨颈的生物力学强度的影响。[方法]取新鲜狗股骨头21对,随机分为3组,每组7对,第1组随机选取一侧,行单纯髓芯减压;第2组随机选取一侧行髓芯减压+钛合金支撑架植入;第3组随机选取其中一侧行单纯髓芯减压,另一侧行髓芯减压+钛合金支撑架植入,治疗模型建立后,在AGS-10KND万能生物材料实验机上进行力学性能测试,观察3组标本股骨颈的力学刚度及强度,对实验结果标准化后行统计学分析。[结果]正常股骨颈,单纯髓芯减压组与髓芯减压+钛合金支撑架植入组股骨颈的刚度及最大负载在统计学上有显著差异。[结论]单纯髓芯减压会降低股骨颈的力学强度,增加术后发生股骨颈骨折的可能,钛合金支撑架植入显著增加了股骨颈的力学强度,对髓芯减压后股骨颈骨折的发生有预防作用。
[ Objective] To study the biomechanics of the femoral neck on the treatment of ischemic of the femoral head in dogs using titanium alloy cage, single core decompression and normal. [ Method] Twenty-one couples of fresh moist dogs' femur specimens were divided randomly into 3 groups. In 2 groups, biomechanical test of the femoral neck was made on one of every couple in different conditions: core decompression, core decompression with titanium alloy cage implanting. The other of every couple was tested as normal control. In the third group, one of every couple was tested in condition of core decompression with titanium alloy cage implanting. The other of every couple was tested in condition of core decompression. [ Result] The stiffness and the max load of the single core decompression were less than that of the normal control. The max load of femoral head after titanium alloy cage implanting were lager than that of the core decompression or the normal control. [ Conclusion ] The titanium alloy cage implanting could increase mechanical support for femoral neck significantly and prevent femoral neck broken after manipulation.
出处
《中国矫形外科杂志》
CAS
CSCD
北大核心
2007年第17期1334-1336,共3页
Orthopedic Journal of China
基金
国家自然科学基金资助项目(30170945)