基于准质心法的采集呼吸信息方法

Acquiring respiratory information on the use of quasi-centroid method
下载PDF
导出
摘要 本文提出准质心方法采集呼吸运动信息。利用超声对两个体模成像,将视频流分解为单帧图像,选择感兴趣区域和图像阈值处理,然后利用准质心方法获取运动组织的呼吸信号。将该方法与相关系数法作比较,包括信号周期、相位和相关分析比较,以此来验证该方法用于信号采集的可行性。结果显示,不同呼吸周期下,两种方法得到的周期误差小于3%,相位差异小于0.08 s,相关分析显示两种方法存在正相关。因此准质心方法也适用于呼吸信号采集。 A method termed as quasi - centroid (QC) proposed to acquire information of respiratory motion.Two phantoms were imaging with B- mode ultrasound unit. Video stream was decomposed into individual image frames. After selection of region of interest and threshold method, phantoms contour derived from threshold were used for QC.A comparison of correlation coefficient (CC) method and QC in period, phage, and correlation analysis were employed to validate the feasibility of acquiring respiratory information of QC. In various environment two phantoms moved, period difference and phase difference for irregular motion are less than 3 %, 0.08s, respectively.And correlation test also indicates good positive correlation. Thereby Qc is also feasible to acquiring respiratory information.
出处 《医疗装备》 2007年第10期6-8,共3页 Medical Equipment
关键词 呼吸运动 准质心法 相关系数法 Respiratory Motion Quasi- centroid Method Correlation Coefficient method
  • 相关文献

参考文献5

  • 1Vedam SS,Keall PJ,Kini VR,et al Acquiring a four-dimensional computed tomography dataset using an external respiratory signal[J].Phys Med Biol,2003,48(1):45-62.
  • 2Shirato H,Shimizu S,Kitamura et al Four-dimensional radiotherapy treatment planning and fluoroscopic real-time tumor tracking for moving tumor[J].Int J Radiat Oncol Biol Phys,2000,48(2):435-442.
  • 3Xu Q,and Hamilton RJ A novel respiratory detection method based on automated analysis of ultrasound diaphragm video[J].Med Phys,2006,33(4):916-921.
  • 4Fuss M,Salter BJ,Cavanaugh SX,et al Daily ultrasound-based image-guided targeting for radiotherapy of upper abdominal malignancies[J].Int J Radiat Oncol Biol Phys,2004,59(4):1245-1256.
  • 5习勉,刘孟忠,邓小武,刘慧,黄晓延,张黎,李巧巧,胡永红,蔡玲,崔念基.应用4D-CT技术确定肝癌内靶体积及相关剂量学研究[J].癌症,2007,26(1):1-8. 被引量:13

二级参考文献22

  • 1刘孟忠,王修身,蔡玲,古模发,刘慧,李群,崔念基,张亚奇,李国辉,李锦清.外照射配合肝动脉栓塞治疗不能手术切除的原发性肝癌的临床研究[J].癌症,2005,24(1):82-86. 被引量:8
  • 2Underberg R W, Lagerwaard F J, Slotman B J, et al. Use of maximum intensity projections (MIP) for target volume generation in 4DCT scans for lung cancer [J]. Int J Radiat Oncol Biol Phys, 2005,63 ( 1 ) : 253-260.
  • 3Engelsman M, Rietzel E, Kooy H M. Four-dimensional proton treatment planning for lung tumors [J]. Int J Radiat Oncol Biol Phys, 2006,64(5) : 1589-1595.
  • 4Wagman R, Yorke E, Ford E, et al. Respiratory gating for liver tumors: use in dose escalation [J]. Int J Radiat Oncol Biol Phys, 2003,55 (3) : 659-668.
  • 5Ten Haken R K, Martel M K, Kessler M L, et al. Use of Veff and iso-NTCP in the implementation of dose escalation protocols [J].Int J Radiat Oncol Biol Phys, 1993,27(3):689-695.
  • 6Lawrence T S, Ten Haken R K, Kessler M L, et al. The use of 3-D dose volume analysis to predict radiation hepatitis[J]. Int J Radiat Oncol Biol Phys, 1992,23(4) :781-788.
  • 7Dawson L A, Normolle D, Baiter J M, et al. Analysis of radiation induced liver disease using the Lyman NTCP model[J]. Int J Radiat Oncol Biol Phys, 2002,53(4):810-821.
  • 8Xu Z Y, Liang S X, Zhu J, et al. Prediction of radiatoninduced liver disease by lyman normal-tissue complication probability model in three-dimensional conformal radiation therapy for primary liver carcinoma [J]. Int J Radiat Oncol Biol Phys, 2006,65( 1 ) : 189-195.
  • 9Baiter J M, Ten Haken R K, Lawrence T S, et al.Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing [J]. Int J Radiat Oncol Biol Phys, 1996,36 ( 1 ) : 167-174.
  • 10Bussels B, Goethals L, Feron M, et al. Respiration-induced movement of the upper abdominal organs: a pitfall for the three-dimensional conformal radiation treatment of pancreatic cancer [J]. Radiother Oncol, 2003,68( 1 ) :69-74.

共引文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部