期刊文献+

猪全血细胞质控物的研究及应用 被引量:1

Study and Application of the QC Material of the Whole Blood Cells of Pig
下载PDF
导出
摘要 目的:猪血制备全血细胞质控物,用于血液学检验质量控制。方法:新鲜抗凝猪全血分离出各种细胞成分,分别用固定剂固定,然后按一定比例混合,防腐防霉处理、定值、分装、保存。结果:制备的全血质控物外观鲜红,显微镜下的血小板、红细胞(BBC)、白细胞(WBC)形态、染色性等特征与新鲜人血相似,无传染性,在sysmex KX-21、A- bout CD-3200、AC920血液分析上能进行多参数检测,瓶间WBC分别为2.6%、3.1%和2.8%,PLT分别为6.3%、6.0%、5.8%,RBC分别为1.5%、1.3%、1.7%,Hb分别为1.6%、1.5%、1.6%,2℃~8℃可稳定6个月,开瓶后可稳定15 d,质控物在sysmex KX-21的测定值与靶值比较,DI值均<1.0。结论:该质控物准确、稳定、无传染性,能满足临床血液学检验质控要求。 Objective The QC material of whole blood cells preparation by the blood of pig,to be used for quality control of the hematology test. Methods To separate the fresh anticoagulant whole blood of pig into diversified cell components which can be fixed by bend, and then let these components fix by some proportion, at last to make antiseptic disposal, to sure the value, to keep by disjoined. Results The QC material of whole blood is red and put it under the microscope, you will find the shape and chromaticity of platelets,red cells and white cells is similar with the fresh blood of the human being. The same QC material of whole blood was divided by several bottles have be measured by sysmex KX-21 ,About CD-3200 and AC920 about many param- eters,the WBC were 2.6% ,3.1% and 2.8% ;PLT were 6.3% ,6.0% and 5.8% ;RBC were 1.5%, 1.3% and 1.7% ;Hb were 1.6% ,1.5% and 1.6% ,the DI data was smaller than 0.1 between measurement and standard by sysmex KX-21. It will be steady for six months under 2℃-8℃ and fifteen days if open the bottles. Conclusion This QC material without infectivity is exact, steady and satisfy, the QC needs of clinical hematology test.
出处 《实用医技杂志》 2007年第23期3132-3133,共2页 Journal of Practical Medical Techniques
关键词 质控物 全血细胞 质量控制 QC material Whole blood cells Quality control
  • 相关文献

同被引文献11

  • 1向甄,杨继庆,文峻.红细胞变形性的影响因素及测量方法[J].中国医学物理学杂志,2006,23(6):423-426. 被引量:14
  • 2中国合格评定国家认可委员会.CNAS-GL03能力验证样品均匀性和稳定性评价指南[S]北京:中国合格评定国家认可委员会,2006.
  • 3Serro AP,Colao R,Saramago B. Adhesion forces in liquid media:effect of surface topography and wettability[J].Journal of Colloid and Interface Science,2008,(02):573-579.doi:10.1016/j.jcis.2008.05.032.
  • 4Svetina S,Ziherl P. Morphology of small aggregates of red blood cells[J].Bioelectrochemistry,2008,(02):84-91.doi:10.1016/j.bioelechem.2007.12.002.
  • 5Chambers E,Mitragotri S. Long circulating nanoparticles via adhesion on red blood cells:mechanism and extended circulation[J].Experimental Biology and Medicine(Maywood),2007,(07):958-966.
  • 6Glenister FK,Fernandez KM,Kats LM. Functional alteration of red blood cells by a megadalton protein of plasmodium falciparum[J].Blood,2009,(04):919-928.doi:10.1182/blood-2008-05-157735.
  • 7Evans J,Gratzer W,Mohandas N. Fluctuations of the red blood cell membrane:relation to mechanical properties and lack of ATP dependence[J].Biophysical Journal,2008,(10):4134-4144.doi:10.1529/biophysj.107.117952.
  • 8Musielak M. Red blood cell-deformability measurement:review of techniques[J].Clinical Hemorheology and Microcirculation,2009,(01):47-64.
  • 9熊立凡;刘成玉.临床检验基础[M]北京:人民卫生出版社,20079-14.
  • 10赵捷,王彩云.尿液有形成分保存方法的研究[J].河北医科大学学报,2003,24(6):367-368. 被引量:1

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部