期刊文献+

氧化钨pH电化学传感器的H^+响应行为研究

Study of the H^+ Response Behavior of WO_3-Based pH Electrochemical Sensor
下载PDF
导出
摘要 采用WO3作为H+响应敏感材料,制备了WO3pH电化学传感器,对传感器的H+响应行为,如响应范围、响应灵敏度、响应时间等影响因素进行了分析。制备WO3工作电极的热处理温度对传感器的H+响应灵敏度与速度有影响,H+响应速度与溶液pH和温度有关,溶液中Na+,K+,F-对H+响应行为没有影响,但是受到I-,NO3-的干扰。 A pH eletrochemical sensor was prepared with a WO3-based working electrode, and the H^+ response behavior of the sensor was investigated, such as pH response range, response sensitivity, response time. The response behavior is mostly effected by the heat-treatment process of the WO3 electrode, while it is also effected by the solution temperature and pH value. The influence of interfered ions was also studied. The result shows that the response of the eletrochemical sensor is not interfered by Na^+ ,K^+ ,F^- ,but by I^- , NO^-3.
出处 《微纳电子技术》 CAS 2007年第7期397-399,共3页 Micronanoelectronic Technology
基金 广东省自然科学基金(05300370) 华南理工大学自然科学青年基金资助项目(B01-E5050840)
关键词 电化学传感器 氧化钨 H^+ 响应 eletrochemical sensor WO3 H^+ response
  • 相关文献

参考文献3

  • 1SHUK P, RAMANUJACHARY K V. New metal-oxide-type pH sensors[J]. Solid State Lonies, 1996, 86-88(2): 1115- 1120.
  • 2LI Q W, LUO G A, SHU Y Qo Response of nano sized cobalt oxide eleetrdes as pH sensors[J]. Analytical Chimiea Acta, 2000, 409(1-2):137-142.
  • 3YAMAMOTO K, SHI G Y, ZHOU T S, et al. Solid-state pH ultramicrosensor based on a tungstic oxide film fabricated on a tungsten nanoelectrode and its application to the study of endothelial cells[J].Analytica Chimica Acta, 2003, 480: 109- 117.