B-FFE序列流动液体信号强度影响因素实验研究 被引量:1

Study of the Factors of Signal Intensity on B-FFE
下载PDF
导出
摘要 目的:实验研究平衡扰相梯度回波(B-FFE)序列影响流动液体信号强度的因素。材料与方法:采用直径2mm(管内径D)的塑胶管,用MR专用高压注射器不同注射流量下(0.5-2.5ml/s)注入纯净水,用飞利浦Intera Achieva1.5T MRI扫描仪B-FFE序列扫描,观察图像信号变化;再用细线结扎出不同管径(1/4-1/2D),Q-FLOW软件测量其信号强度。结果:(1)以0.5、1.0、1.5、2.0、2.5ml/s的不同流量注射纯净水后,信号强度与流量无关,均呈均匀高信号。(2)以不同程度结扎管径,结扎后信号强度下降,但随着结扎程度和次数增加,信号强度下降程度显著、范围延长。结论:在B-FFE序列中,流动流体信号强度与流动速度无关,而管径狭窄是信号强度降低的主要因素。 Purpose: To experimental study the effecting factors of the signal intensity (SI) of flow fluid on balance - fast field echo (B - FFE) sequence. Materials and Methods: B- FEE sequences of different velocity water (0.5- 2.5ml/s) of a plastic tube (2mm in diameter) through MRI injector were obtained by using Pifilips Intera Achieva 1 .ST MRI system, and the changes of signal intensity (SI) were observed. Then plastic tube was ligated by thin string to different diameters (1/2- 1/4 diameter), SI was measured by Q- flow software. Results: (1) The SI was independent On the fluid flow velocity, which were all high SI. (2) Decrease of SI was found after the plastic tube ligated. Tire more the tube was ligated in degree, the more the SI decreased and extended. Conclusion: The SI is independent on fluid flow velocity, and dependent on tube narrow mutually.
出处 《现代医用影像学》 2007年第3期100-103,共4页 Modern Medical Imageology
关键词 血流信号强度 B-FFE MRI Blood SI B-FFE MRI
  • 相关文献

参考文献6

二级参考文献42

  • 1孙福成,许锋,季福绥,何青,吴军,林颖,杨玉珍,沈月莉,王树华.球囊扩张肾动脉成形术及肾动脉支架治疗肾动脉狭窄[J].中国介入心脏病学杂志,2004,12(3):188-188. 被引量:3
  • 2朱晓黎,沈天真,陈星荣.MR相位对比电影法在正常颈部及颅内大血管血流测量中的应用[J].中国医学影像技术,2005,21(3):365-368. 被引量:17
  • 3史玉泉.实用神经病学(第2版)[M].上海:上海科学技术出版社,1994.736.
  • 4James CC, Orlando S, Jeff B, et al. Cine MR angiography of the heart with segmented true fast imaging with steady-state precession[J]. Radiology, 2001,219(3) :828- 834.
  • 5李刊成.心血管磁共振诊断学[M].北京:人民卫生出版社,1997,1..
  • 6Tan SP, Bux SI, Kumar G, et al. Evaluation of live renal donors with three-dimensional contrast-enhanced magnetic resonance angiography in comparison to catheter angiography. Transplant Proc, 2004, 36(7):1914
  • 7Giessing M, Kroencke TJ, Taupitz M, et al. Gadolinium-enhanced three-dimensional magnetic resonance angiography versus conventional digital subtraction angiography: which modality is superior in evaluating living kidney donors? Transplantation,2003,76(6):1000
  • 8Kaul MG, Stork A, Bansmann PM, et al. Evaluation of Balanced Steady-State Free Precession (TrueFISP) and K-space Segmented Gradient Echo Sequences for 3D Coronary MR Angiography with Navigator Gating at 3 Tesla. Rofo, 2004,176(11):1560
  • 9Tan KT, Van Beek EJR, Brown PWG, et al. Magnetic resonance angiography for the diagnosis of renal artery stenosis: a meta-analysis. Clinical Radiology, 2002,57:617
  • 10Arlart IP, Bongartz GM. Mgnetic Resonance Angiography. 2nd Revised Edition, Springer-Verlag Berlin Heidelberg, 2002,297

共引文献30

同被引文献10

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部