两种眼压计对环扎术后眼压测量的比较

Comparison of Two Different Tonometers Following Encircling Operation
下载PDF
导出
摘要 目的弄清非接触式眼压计(Non-ContactTonometer,NCT)和Schitz眼压计对环扎术后的眼压测量有何差异。方法用NCT和Schitz眼压计对57只视网膜脱离环扎加压术后的眼进行眼压测量。患者健眼亦57只作对照组。结果手术组的NCT值为1.902±0.664kPa;Schitz值为0.943±0.464kPa;对照组NCT值为2.381±0.620kPa;Schitz值为1.942±0.450kPa;手术组的(NCT-Schitz)值为0.958±0.435kPa(P<0.001);对照组(NCT-Schitz)值为0.439±0.491kPa(P<0.001)。结论在环扎加压术后的眼压测量中NCT比Schitz眼压计准确。 ObjectiveTo compare non contact tonometer (NCT) with Schiotz tonometer applied to retinal detachment eyes after encircling procedure.Methods57 cases in 57 eyes with retinal detachment were measured with NCT and Schiotz tonometer 3 4 days after encircling operation.The healthy fellow eyes served as the control group. Results The mean NCT value of treated group was 1.902±0.664kPa and the mean Schiotz value was 0.943±0.464kPa.The mean NCT value of control group was 2.381±0.620kPa,the mean Schiotz value was 1942±0.450kPa.The NCT Schiotz of treated group was 0.958±0.435kPa, while that of control group was 0.439±0.491kPa.ConclusionNCT is more accurate than Schiotz tonometer in the measurement of IOP in eyes after encircling procedure
作者 田小波 张晰
出处 《眼科研究》 CSCD 1997年第2期117-118,共2页 Chinese Ophthalmic Research
关键词 环扎加压 非接触眼压计 眼压测量 ncircling procedure non contact tonometer tonometry