摘要
目的弄清非接触式眼压计(Non-ContactTonometer,NCT)和Schitz眼压计对环扎术后的眼压测量有何差异。方法用NCT和Schitz眼压计对57只视网膜脱离环扎加压术后的眼进行眼压测量。患者健眼亦57只作对照组。结果手术组的NCT值为1.902±0.664kPa;Schitz值为0.943±0.464kPa;对照组NCT值为2.381±0.620kPa;Schitz值为1.942±0.450kPa;手术组的(NCT-Schitz)值为0.958±0.435kPa(P<0.001);对照组(NCT-Schitz)值为0.439±0.491kPa(P<0.001)。结论在环扎加压术后的眼压测量中NCT比Schitz眼压计准确。
ObjectiveTo compare non contact tonometer (NCT) with Schiotz tonometer applied to retinal detachment eyes after encircling procedure.Methods57 cases in 57 eyes with retinal detachment were measured with NCT and Schiotz tonometer 3 4 days after encircling operation.The healthy fellow eyes served as the control group. Results The mean NCT value of treated group was 1.902±0.664kPa and the mean Schiotz value was 0.943±0.464kPa.The mean NCT value of control group was 2.381±0.620kPa,the mean Schiotz value was 1942±0.450kPa.The NCT Schiotz of treated group was 0.958±0.435kPa, while that of control group was 0.439±0.491kPa.ConclusionNCT is more accurate than Schiotz tonometer in the measurement of IOP in eyes after encircling procedure
出处
《眼科研究》
CSCD
1997年第2期117-118,共2页
Chinese Ophthalmic Research
关键词
环扎加压
非接触眼压计
眼压测量
ncircling procedure non contact tonometer tonometry