影响计算机X线摄影图像质量因素的临床分析 被引量:2

Clinical Analysis of the Influential Factors of the Imaging Quality in Computed Radiography
下载PDF
导出
摘要 目的:探讨影响计算机X线摄影(CR)图像质量的因素及优化图像质量的方法。方法:分析2000例不同部位计算机X线摄影成像的技术参数、伪影、IP板使用方式及后处理方法,并与计算机X线摄影的图像质量标准作对照研究,分析影响图像质量的各相关因素。结果:2000例中甲级片1720张,乙级片182张,丙级片98张。乙级与丙级片共280张,其中因摄影技术不当所致141张,伪影引起65张,后处理不当74张。结论:计算机X线摄影图像质量受多种因素影响,合理设置参数及采取各种针对措施能有效提高图像质量。 Objective:To inspect the influential factors of the imaging quality in computed radiography and find the methods to improve the imaging quality. Methods:The imaging parameters,artifacts, the application of IP plate and the postprocessing techniques of 2000 pictures from various parts of human body,were studied,and compared with images from X- ray computed radiography. The relationship bwtween the influential factors and imaging quality was researched. Results: In all of 2000 pictures,the count of superior,medium and inferior imaging quality pictures were 1720,182 and 98 respectively. The influential factors of image quality include imaging parameters,artifact and the posprocessing techniques. Conclusion: The imaging quality of computed radiography was affected by various factors. The optimized imaging parameters and the application of various corresponding measures could improve the imaging quality.
出处 《放射学实践》 2007年第9期985-987,共3页 Radiologic Practice
关键词 体层摄影术 X线计算机 诊断 计算机辅助 图像处理 计算机辅助 Tomography, X-ray computed Diagnosis, computer-assisted Image processing, computer-assisted
  • 相关文献

参考文献5

二级参考文献10

  • 1全国放射科QA,QC学术研讨会纪要[J].中华放射学杂志,1993,27(2):134-138. 被引量:134
  • 2Pieter V,Piet D,Emile S.Optimizing Computed Radiography Imaging Performance[R].Washington-Georgetown University, 1997.
  • 3Cesar LJ,Schueler BA,Zink FE,Daly TR,Taubel JP,Jorgenson LL.Artifacts found in computed radiography [J] .The British Journal of Radiology,2001,74: 195 - 202.
  • 4Solomon SL,Jost RG,Glazer HS,Sagel SS,Anderson DJ,M olina PL .Aratfacts in computed radiography[J].AJR, 1991,157:181 - 5.
  • 5Ostemann JW ,Prokop M ,Schaefer CM ,Galanshi M .Hardware and software artifacts in storage phosphor radiography[J]. Radiographic, 1991,11 :795 - 805.
  • 6AAPM Report.Acceptance testing and quality control of photostimulable storage phosphor imaging systems. Report of the Task Group No.10.diagnostic x- ray imaging committee,JA Seibert,chair.NewYork,NY :Am Assoc Physicists M ed;in press.
  • 7燕树林 范英.放射技术的质量管理与组织管理[M].,1999..
  • 8张泽宝 等.医学影像物理学[M].北京:人民卫生出版社,2001.108-158.
  • 9Chotas HG,Floyd CE Jr,Ravin CE.Memory artifact related to selenium-based digital radiography systems.Radiology,1997,203:881-883.
  • 10Cesar LJ,Schueler BA,Zink FE,et al.Artefacts found in computed radiography(Review).Br J Radiol,2001,74:195-202.

共引文献55

同被引文献9

引证文献2

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部