图像引导自主呼吸控制下肝脏肿瘤计划靶区外放研究 被引量:3

Design of planning target volume margin using an active breathing control and Varian image-guided radiotherapy(IGRT) system in unresectable liver tumor
原文传递
导出
摘要 目的 图像引导自主呼吸控制(ABC)下肝脏肿瘤适形放疗计划靶体积(刚)外放范围。方法 应用瓦里安公司图像引导放疗(IGRT)系统,对13例经肝动脉栓塞化疗(TACE)的患者,在kV透视测量ABC下碘油影像的稳定性后,以锥形束CT(CBCT)联合ABC技术于每次治疗前采集容积CT图像10次/人。根据碘油影像进行kV-CBCT和计划CT的3D-3D配准,并以此校正摆位误差,设定修正摆位的阈值为3mm。治疗结束后再次采集kV-CBCT和计划CT3D-3D配准验证。结果 ABC kV透视模式下碘油几乎不存在运动。根据stroom公式M=2.0∑+0.7α和测量结果进行计算,治疗前不加修正和使用ABC技术时肝脏治疗的PTV外放x.y、z轴分别为4.4、5.3、7.8mm。采用以上技术进行在线校正时,PTV外放x,y、z轴分别为2.5、2.6、3.9mm。结论 应用ABC技术的肝脏适形放疗,建议PTV外放x、y、z轴分别为5、6、8mm,应用ABC技术结合图像引导适形放疗系统在线治疗前校正时,PTV外放x,y、z轴分别为3、3、4mm。 Objective To-define the planning target voiume(PTV) margin with an active breathing control (ABC) and the Varian image-guided radiotherapy (IGRT) system. Methods Thirteen patients with liver cancer were treated with radiotherapy from May 2006 to September 2006. Prior to radiotherapy, all patients had undergone transarterial chemoembolization (TACE) by infusing a mixture of iodized oil contrast medium and chemotherapeutic agents, kV fluoroscopy was used to measure the potential motion of lipiedol spot positions during ABC breath-holds. ABC was used for planning CT scan and radiation delivery with the breath held at the same phase of the respiratory cycle (near end-exhalation). Cone beam CT (CBCT) was taken using Varian IGRT system,which was then compared online with planning CT using a 3D-3D matching tool . Analysis relied on lipiodol spots on planning CT and CBCT mannully. The treatment table was moved to produce acceptable setup before treatement delivery. Repeated CBCT image and another analysis were obtained after irradiation. Results No motion of the intrahepatic tumor was observed on fluoroscopy during ABC breath-holds. The estimated required PTV margins, calculated according to the stroom formula, were 4.4 mm, 5.3 mm and 7.8 mm in the x, y and z axis directions before radiotherapy. The corresponding parameters were 2.5m, 2.6 mm and 3.9 mm after radiotherapy. Conclusions We have adopted a PTV margin of 5 mm,6 mm and 8 nun in the x, y and z axis directions with ABC, and 3,3 and 4 mm with ABC and on-line kilovohage CBCT.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2007年第5期377-380,共4页 Chinese Journal of Radiation Oncology
关键词 肝脏肿瘤/三维适形放疗 图像引导放疗系统 锥形束CT 自主呼吸控制 计划靶体积 Liver neoplasms Image guided radiation therapy system Cone-beam CT Active breathing control Planning target volume
  • 相关文献

参考文献19

  • 1Wong JW, Sharpe MB, Jaffray DA, et al. The use of active breathing control (ABC) to reduce margin for breathing motion. Int J Radiat Oncol Biol Phys, 1999,44:911-919.
  • 2戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 3ICRU50. Prescribing, Recording and Reporting Photon Bream Therapy. Bethesda:ICRU, 1993.
  • 4Remeijer P, Geerlof E, Ploeger L, et al. 3-D portal image analysis in clinical practice: an evaluation of 2-D and 3-D analysis techniques as applied to 30 prostate cancer patients. Int J Radiat Oncol Biol Phys ,2000,46 : 1281-1290.
  • 5Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report. Radiother Oncol,2002, 64:75-83.
  • 6Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys, 1999,43:905-919.
  • 7Xing L, Thorndyke B, Schreibmann E, et al. Overview of imageguided radiation therapy. Med Dosim,2006 ,31:91-112.
  • 8Jaffray DA, Siewerdsen JH. Cone-beam computed tomography with a flat-panel imager: initial performance characterization. Med Phys, 2000,27:1311-1323.
  • 9Jaffray DA, Siewerdsen JH, Wong JW, et al. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys,2002,53 : 1337-1349.
  • 10Robertson JM, Lawrence TS, Andrews JC, et al. Long-term results of hepatic artery fluorodeoxyuridine and conformal radiation therapy for primary hepatobiliary cancers. Int J Radiat Oncol Biol Phys, 1995,37:325-330.

二级参考文献18

  • 1.调强适形放射治疗[A].胡逸民.肿瘤放射物理学[C].北京:原子能出版社,1999.538-612.
  • 2Hanley J,Debois MM,Mah D,et al.Deep inspiration breath-hold technique for lung tumors:the potential value of target immobilization and reduced lung density in dose escalation.Int J Radiat Oncol Biol Phys,1999,45:603-611.
  • 3Ramsey CR,Scaperoth D,Arwcod D,et al.Clinical efficacy of respiratory gated conformal radiation therapy.Med Dosim,1999,24:115-119.
  • 4Shirato H,Shimizu S,Kitamura K,et al.Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.Int J Radiat Oncol Biol Phys,2000,48:435-442.
  • 5Keall PJ,Chen GTY,Joshi S,et al.Time-the fourth dimension in radiotherapy (ASTRO panel discussion).Int J Radiat Oncol Biol Phys,2003,57:S8-S9.
  • 6Yu CX,Jaffray DA,Wong JW.The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation.Phys Med Biol,1998,43:91-104.
  • 7de Neve W,Heuvel F,De Beukeleer M,et al.Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch.Radiother Oncol,1992,24:45-54.
  • 8Hu YM,DM JR,Hu B.The precision and accuracy analysis of target position in X-ray stereotactic radiotherapy for extracranial lesions.In:Internation symposium of stereotactic radiosurgery (SRS/SRT).Taejon:Chungnam National University Hospital,1997.
  • 9Yan D,Vicini F,Wong J,et al.Adaptive radiation therapy.Phys Med Biol,1997,42:4123-4132.
  • 10Birkner M,Yan D,Alber M,et al.Adapting inverse planning to patient and organ geometrical variation algorithm and implementation.Med Phys,2003,30:2822-2831.

共引文献183

同被引文献69

引证文献3

二级引证文献43

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部