摘要 目的探讨使用电子射野影像装置(EPID)对放疗设备MLC的灯光野与射线野一致性进行调整,以快速的方法,更可以达到节省人力物力和时间。方法以10CM×10CM、20CM×20CM、30CM×30CM照射野为参考条件,分别以Elekta-iViewGTTM系统跟踪测量每个实验点灯光野与射线野一致性的改变并分析误差程度,对MLC的叶片位置进行参数设定,最后达到一致。结果iViewGTTM系统得出的数据和剂量胶片法分析取得的灯光野与射线野的一致性的参数之间误差比较不大于0.7mm。结论可以利用电子射野影像装置(EPID)验证放疗设备的灯光野与射线野的一致性,而且是较准确、有效、可靠和操作简单快捷直观以及数字化程度高,替代传统的测验手段。 Objective To study using electronic: imaging device to adjust light and radiation field angruenee of MLC,with rapid wetheds to save many a lot of human being and time.Methods With field 10 10CM×10CM、20CM×20CM、30CM×30CM for reference, We use system Elekta-iViewGTTMto measure every port's difference of the light field and the radiation Field, then Set the parameter of the MLC leave's position to accord the light field with the radiation Field.Result The data error of the consistency of the light field and the radiation Field out from the iViewGT^TM system is less than 0.7mm compare with the fihn method.Conclusions We can use the EPID to validate the consistency of the light field and the radiation Field. It nicety, effieieney, reliable method that can substitute the traditional measurement.
出处 《中国医学装备》 2007年第9期44-46,共3页 China Medical Equipment
关键词 iViewGT 射线野 误差 iViewGT,radiation field,error
  • 相关文献

参考文献2

二级参考文献17

  • 1沈瑜.临床放射生物学研究进展[J].国外医学(放射医学核医学分册),1996,20(1):35-38. 被引量:2
  • 2胡逸民,肿瘤放射物理学,1999年,613页
  • 3胡逸民 杨定宇.肿瘤放射治疗技术[M].北京:中国协和医科大学、北京医科大学联合出版社,1995..
  • 4Edward S, Sternick,Ph D.Intensity modulated radiation therapy. Proceedings of the internationmal conference on medical imaging[J].Medical physics and precision radiation therapy, 1939,10:32~36.
  • 5Hunt MA, Hsiung CY,Spirou SV,et al. Evaluation of concave dose distributions created using an inverse planning system[J].Int J Radiat Oncol Biol. Phys.,2000,48(Suppl 3):138.
  • 6Nutting CM, Rowbottom C,Cosgrove UP,et al. Evaluation of intensity modulated radiotherapy for tumors of the pelvis and head and neck and thorax optimization and clinical implementation [J].Int J Radiat Oncol Biol. Phys.,2000,48(Suppl 3):139.
  • 7Yu CX.Intensity-modulated are therapy with dynamic multileaf collimation:an alternative to tomotherapy[J].Phys. Med. Biol.,1995,40:1435~1449.
  • 8陈炳桓.立体定向放射[M].北京:北京出版社,1994.1-113.
  • 9L eksell L. Stereotaxis and radio surgery:an operative system[M].Spring field,IL:Charles C.Thomas,1971,1~20.
  • 10L unsford L D, Flickinger J, Lindner G,et al. Stereotactic radiosurgery of the brain using the first united states 201 cobalt 60source gamma knife[J].Neurosurgery, 1989,24:151~158.