摘要
采用王水分解试样-原子荧光光谱法(AFS)测定粉煤灰中的砷、汞,方法回收率分别为98.5%~102.0%、85.0%~105.0%,相对偏差分别为1.9%~4.8%、0.6%~2.4%.方法具有快速、简便、灵敏等优点,对探讨粉煤灰在农业生态环境方面的综合利用颇有价值.
The sample was dissolved in aqua vegia; AFS method was used to determine arsenic and mercury in coal fly ash. The ralative stand deviations of them are 1.9%-4.8%,0. 6%-2.4%. The recovery of them is 98. 5% - 102. 0%, 85. 0% - 105. 0% respectively. The method is simple, quick and high sensitivity. The method has been applied to comprehensive utilization of coal fly ash in monitor agriculture ecology environment.
出处
《分析测试技术与仪器》
CAS
2007年第3期219-221,共3页
Analysis and Testing Technology and Instruments
关键词
原子荧光光谱法
粉煤灰
砷
汞
atomic fluorescence spectrometry
coal fly ash
arsenic
mercury