期刊文献+

废水COD快速开管测定方法研究 被引量:1

Studies on COD by Quick opening-tube measurement method
下载PDF
导出
摘要 目的:为探讨COD的快速测定方法。方法:采用硫酸锰代替硫酸银作催化剂,研究不同消解温度、时间、酸度和催化剂用量对COD测定的影响。结果:开管法快速测定COD的最佳操作条件为:消解温度165℃,消解时间12min,反应体系的酸度10.8 mol/L,硫酸锰用量10 g/L。结论:与标准法相比,具有简便、快速,分析成本低、准确度和灵敏度高的优点。可以代替标准法广泛用于各种废水COD的测定。 In order to discuss COD the rapid determination method, Using MnSO4 Method as catalyst instead of Ag2SO4,studies is for COD influence in different temperature, time, acidity and catalyst dosage. The optimum determining conditions are as follows: decomposition temperature 165℃, ecomposition time 12 min, acidity 10.8 mol/L, MnSO4dosage 10 g/L. Compared with the standard methods, the method has the advantages of simplicity, quickness, reduced the analysis cost, high accuracy and sensitivity. Many water samples can he dissociated with the method and it fits for the determination of bulky water samples.
作者 娄晶晶
出处 《内蒙古石油化工》 CAS 2007年第9期10-12,共3页 Inner Mongolia Petrochemical Industry
关键词 COD快速测定 催化剂 硫酸锰 开管法 COD rapid determination, catalyst manganese sulfate opening--tube method
  • 相关文献

参考文献3

二级参考文献7

  • 1国家环保局.水和废水监测分析方法[M].北京:中国环境科学出版社,1989.260-263.
  • 2中国医学科学院卫生研究所.水质分析法[M].北京:人民卫生出版社,1972.192-195.
  • 3奚旦立 孙玉生 等.环境监测[M].北京:化学工业出版社,1991.214-232.
  • 4国家环保局科技处.我国几种工业废水治理技术研究(第三册).化学工业出版社,1998
  • 5国际标准化组织 汇编,刘振庄.水与大气质量分析方法国际标准[M]中国环境科学出版社,1990.
  • 6《水和废水监测分析方法指南》编委会,.水和废水监测分析方法指南[M]中国环境科学出版社,1989.
  • 7张莘民,高明慧.化学需氧量的分析[J].环境化学,1989,8(2):62-69. 被引量:15