摘要 与光学透镜和CCD的耦合方法相比,光锥和CCD的耦合器件具有体积小、重量轻、耦合效率高的特点。分析了光锥和CCD的耦合效率对整个成像系统信噪比的影响,阐述了光锥和像增强器、CCD的耦合方法。实验测试结果表明:成像系统的空间分辨力达到4lpixel/mm;对人手的X射线透视结果表明:手部骨骼轮廓清晰,能够对其病变做出准确诊断。 Compared to a lens-coupled CCD camera system,the CCD camera coupled with optical fiber taper has less cubage,lighter weight and higher collection efficiency. The influence of coupling efficiency of optical fiber taper with CCD on signal-to-noise ratio of image system is analysed. In addition , the coupling method of optical fiber taper coupling with generation image intensifier and CCD is expatiated. Inspection result shows that the resolution of imaging system is 41 pixel/mm and the expeiment for radiographic image of people hand bones indicates that imaging system can be used to medical diagnoses.
出处 《传感器与微系统》 CSCD 北大核心 2007年第9期26-28,共3页 Transducer and Microsystem Technologies
基金 山西省高等学校科技研究开发项目(20041222)
关键词 光锥 X射线像增强器 电荷耦合器件 空间分辨力 optical fiber taper X-ray generation image intensifier CCD spatial resolution
  • 相关文献

参考文献3

二级参考文献4

共引文献75

同被引文献19

  • 1汪丽,田维坚.光锥CCD耦合器件对提高CCD成像分辨率的探讨[J].光电子技术与信息,2004,17(3):21-25. 被引量:3
  • 2Woods A J, Penrose J D. The development of a compactunderwater stereoscopic video camera[J]. SPIE, 1997,3012:92-95.
  • 3Trimble G M. Area reconnaissance, object relocation andclassification using cooperating autonomous underwatervehiclesfj]. SPIE, 1999, 3711:195-202.
  • 4Kubota S,Shimada H. High-speed photo-graphy of under-water sympathetic detonation of explosives[J]. SPIE, 2001,4183:760-770.
  • 5Evans C, Doshir. Tale of two underwater lenses[J].SPIE,1991, 1537:203-214.
  • 6Woods A J, Penrose J D. The development of a compactunderwater stereoscopic video camera[J]. SPIE, 1997,3012:92-95.
  • 7Trimble G M. Area reconnaissance, object relocation andclassification using cooperating autonomous underwatervehiclesfj]. SPIE, 1999, 3711:195-202.
  • 8Kubota S,Shimada H. High-speed photo-graphy of under-water sympathetic detonation of explosives[J]. SPIE, 2001,4183:760-770.
  • 9Evans C, Doshir. Tale of two underwater lenses[J].SPIE,1991, 1537:203-214.
  • 10闫晓梅,王志社.基于光锥耦合的X射线像增强器[J].光学学报,2010,30(5):1478-1482. 被引量:6

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部