摘要
应用单纯形最优化法研究了三种混合试剂体系消除原子吸收分光光度法测定钙时铝的干扰。在所选体系下测定样品,其回收率为93%~105%,结果令人满意。
The three mixed releasing reagent systems of removing interference of aluminium in the atomic absorptive spectrometric determination of calcium were studied with simplex optimization. The recovery of sample under the optimized systems was 93%~105%. The result was satisfactory.
出处
《青岛化工学院学报(自然科学版)》
1997年第1期58-62,共5页
Journal of Qingdao Institute of Chemical Technology(Natural Science Edition)
关键词
单纯形最优化法
原子吸收
分光光度
钙
铝
测定
simplex optimization
atomic absorption spectrometry
calcium
Aluminium