期刊文献+

人工髋关节置换前后股骨及假体的生物力学分析 被引量:11

Biomechanical analysis of prosthesis and femur before and after hip replacement
下载PDF
导出
摘要 目的:研究人工髋关节置换前后股骨及假体的应力分布以及假体设计参数对应力的影响。方法:建立股骨和假体有限元模型,分析在步行峰值关节载荷和主要肌肉载荷作用下,完整股骨和置换后股骨及假体的应力水平。结果:完整股骨中上部内侧受压应力,外侧受张应力,中下部外侧受压应力,内侧受张应力,股骨应力峰值位于中下部;置换后股骨受力总体模式不变,近端应力遮挡显著。随颈干角增加,假体及股骨应力水平降低;柄长对假体应力影响不大,股骨上的应力随柄长增大略有增加。结论:假体设计时可适当增大颈干角,在骨组织条件允许的情况下可适当增加假体柄长,以减轻术后的应力遮挡效应。 Objective: To determine stress distribution of prosthesis and femur before and after hip replacement and the effect of design parameters on the stress. Methods: The finite element models of an intact femur and a femur implanted with a femoral prosthesis were developed and stress analysis were carried out when both joint contact forces and muscle forces were included in the loading. Results: The peak stresses of intact femur were presented at the middle and lower regions of femur with compressive stress in lateral side and tensile stress in medial side, which was on the contrary at the upper femur. The stress pattern of implanted femur was similar to the one of intact femur with the obvious stress shielding at proximal femur. With increasing neck-shaft angle, the stress of femur and prosthesis decreased. The increased length of the stem brought the increased stress of femur slightly, but had little effect on the stress of prosthesis. Conclusions: Neck-shaft angle could be increased properly and stem length could be increased in certain conditions to reduce the stress shielding in implanted femur.
出处 《中国临床解剖学杂志》 CSCD 北大核心 2007年第5期579-582,共4页 Chinese Journal of Clinical Anatomy
关键词 人工髋关节 股骨 有限元 生物力学 应力遮挡 artificial hip joint femur finite element biomechanics stress shielding
  • 相关文献

参考文献10

  • 1Dan D,Germann D,Burki H,et al.Bone loss after total hip arthroplasty[J].RheumatolInt,2006,26(9):792-798.
  • 2Wang CJ,Brown CJ,Yettram AL,et al.Intramedullary nails:some design features of the distal end[J].Med Eng Phys,2003,25(9):789-794.
  • 3Bergmann F,Graichen,Rohlmann,A.Hip joint loading during walking and running,measured in two patients[J].J Biomech,1993,26(8):969-990.
  • 4Gotze C,Tschugunow A,Gotze HG,et al.Long-term results of the metal-cancellous cementless Lubeck total hip arthroplasty:a critical review at12.8 years[J].Arch Orthop Trauma Surg,2006,126(1):28-35
  • 5严世贵,何荣新,陈维善,吴浩波,余世策.全髋关节置换前后股骨应力变化的有限元分析[J].中华骨科杂志,2004,24(9):561-565. 被引量:62
  • 6汪金平,杨天府,钟凤林,殷光义,兰玉平,杨小奇.股骨生物力学特性的有限元分析[J].中华创伤骨科杂志,2005,7(10):931-934. 被引量:61
  • 7Taylor ME,Tanner KE,Freeman MAR,et al.Stress and strain distribution within the intact femur:compression or bending[J]? Med Eng Phys,1996,18(2):122-131.
  • 8Mjoberg B.The theory of early loosening of hip prostheses[J].Orthopedics,1997,20(12):1169-1175.
  • 9Schmidt R,Nowak TE,Mueller L,et al.Osteodensitometry after total hip replacement with uncemented taper-design stem[J].Int Orthop,2004,28(2):74-77.
  • 10Jolles BM,Zangger P,Leyvraz PF.Factors predisposing to dislocation after primary total hip arthroplasty:a multivariate analysis[J].J Arthroplasty,2002,17(3):282-288.

二级参考文献14

  • 1Visuri T, Hietaniemi K. Displaced stress fracture of the femoral shaft: a report of three cases. Mil Med, 1992, 157: 325-327.
  • 2Nikolopoulos KE, Papadakis SA, Kateros KT, et al. Long-term outcome of patients with avascular necrosis, after internal fixation of femoral neck fractures. Injury, 2003, 34: 525-528.
  • 3Schep NW, Heintjes RJ, Martens EP, et al. Retrospective analysis of factors influencing the operative result after percutaneous osteosynthesis of intracapsular femoral neck fractures.Injury, 2004, 35: 1003-1009.
  • 4Selvan VT, Oakley M J, Rangan A, et al. Optimum configuration of cannulated hip screws for the fixation ofintracapsular hip fractures:a biomechanical study. Injury, 2004, 35: 136-141.
  • 5Levi N, Ingles A Jr, Klyver H, et al. Fracture of the femoral neck: optimal screw position and bone density determined by computer tomography. Injury, 1996, 27: 287-289.
  • 6Booth KC, Donaldson TK, Dai QG. Femoral neck fracture fixation: a biomechanical study of two cannulated screw placement techniques. Orthopedics, 1998, 21: 1173-1176.
  • 7Woo SL, Simon BR, Akeson WH, et al. A new approach to the design of internal fixation plates. J BiomedMater Res, 1983,17: 427-439.
  • 8Huiskes R, Chao EY. A survey of finite element analysis in orthopedic biomechanics: the first decade. J Biomech, 1983,16: 385-409.
  • 9King AI. A review of biomechanical methods. J Biomech Eng,1984, 106: 97-104.
  • 10Taylor ME, Tanner KE, Freeman MA, et al. Stress and strain distribution within the intact femur: compression or bending?Med Eng Phys, 1996, 18: 122-131.