期刊文献+

石墨炉原子吸收法直接测定血液中锡的方法研究 被引量:1

Determination of tin in blood by electrothermal atomic absorption spectrometry
下载PDF
导出
摘要 目的:建立测定全血中锡的新方法。方法:应用Triton X-100稀释样品,横向塞曼石墨炉原子吸收光谱法直接测定样品中痕量锡。结果:选定了最佳测量条件,该方法的灵敏度、精密度均能满足检测工作的需要。本法对锡的最低检出量为0.63Pg,回收率在92.3%~99.4%之间。结论:本法简便、快速、准确,灵敏度和精密度高,适用于全血样品中痕量锡的测定。 Objective:To establish a new method for determination of tin in blood. Methods:The samples were diluted in a medium containing Triton X - 100 before being introduced directly into the furnace and then detected by Zeeman GFAAS. Results: Under the best condition, the sensitivity and precision were suitable for the determination while the rate of recovery was controlled between 92. 3 percent to 99. 4 percent. The lowest detection limit of tin was 0. 63 pg. Conclusion:This method is simple, quick, accurate and sensitive. It is a very good method for detecting tin in blood.
出处 《中国卫生检验杂志》 CAS 2007年第9期1603-1604,共2页 Chinese Journal of Health Laboratory Technology
关键词 横向塞曼GFAAS全血 Landscape orentation Zeeman GFAAS Blood Tin
  • 相关文献

参考文献3

  • 1Sandborgh Englund G,Bjorkman L,Bjorkman L,et al. Determination of low levels of total mercury in blood and plasma by cold vapour atomic fluorescence spectrometry[J].Scand J Clin Lab Invest,1998,58(2):155-160.
  • 2Perring L, Basic Dvorzak M. Determination of the different states of mercury in seawater near the Vlora and Durres Bays[J]. Anal Bioanal Chem,2002,374(6):1034-1038.
  • 3孙昕,洪素珍.钨盐涂覆-塞曼无火焰原子吸收法测定肿瘤患者血清中锡[J].光谱学与光谱分析,1995,15(1):79-82. 被引量:3