摘要 建立大体积直接进样离子色谱法测定饮用水中痕量无机消毒副产物溴酸盐和4种常见的阴离子。采用Ion Pac AS9-HC阴离子色谱柱,以8 mmol/LNa2CO3+2 mmol/LNaOH为淋洗液,进样体积240μL,流速为0.25 mL/min,在30 min内可以有效测定溴酸盐。溴酸盐的检出限为1.86μg/L,加标回收率为89.5%~110.2%。 A novel method was developed for the determination of bromate in drinking water by ion chromatography with large volume direct injection. With a high capacity anion exchange column , IonPac AS9-HC column , and 8 mmol/L Na2CO3+2 mmol/LNaOH as eluent, the bromate could be well separated with other anions and determined within 30 mins with a flow rate of 0.25mL/min.A 240 μL sample loop was used. The detection limits for bromate was 1.86 μg/L. The recoveries of the method for bromate was between 89.5%-110.2%.
出处 《净水技术》 CAS 2007年第5期73-75,共3页 Water Purification Technology
基金 国家自然科学基金(20477007) 广州市科技计划(2004J1-C0331) 广东省重大科技专项(2004A3038002)
关键词 离子色谱 直接进样 溴酸盐 阴离子 ion chromatography direct injection bromate anion
  • 相关文献

参考文献4

  • 1Melnick R L. An alternative hypothesis on the role of chemically induced protein droplet nephropathy in renal carcinogenesis [J]. Reguletory Toxieol and Pharmaeol, 1992,16(3):111-125
  • 2Determination of Inorganic Anions in Water by Ion Chromatography [S]. U.S. Environmental Protection Agency Method 300.1, Cincinnati, Ohio, 1997
  • 3Determination of Inorganic Oxyhalide Disinfection By-products in Drinking Water Using Ion Chromatography with the addition of a post-column reagent for trace bromate Analysis [S]. U. S.Environmental Protection Agency Method 317.2 , Cincinnati, Ohio, 2001
  • 4Determination of Bromate in Drinking Waters by Ion Chromatography Inductively Coupled Plasma-Mass Spectrometry [S]. U.S. Environmental Protection Agency Method 321.8, Cincinnati, Ohio, 1997