期刊文献+

镉-铜法测定海水中硝酸盐的过度还原问题 被引量:4

THE EXCESSIVE REDUCTION PROBLEMS IN NITRATE ANALYSIS (Cd-Cu METHOD)
下载PDF
导出
摘要 认为用镉 -铜法测定硝酸盐时还原率低的原因是过度还原 ,并介绍了判断过度还原存在与否的实验方法 ,研究了影响过度还原程度的因素及解决方法。 Usually it is considered that the nitrate reduction column should be reactivated if the reduction efficiency is less than 95 %,because the reduction column has low reduction ability, but this study indicated that the low reduction efficiency may also caused by excessive reduction. A simple experiment method to determine whether excessive reduction has occurred was introduced, and the factors which influenced the excessive reduction degree were discussed.
出处 《海洋科学》 CAS CSCD 北大核心 1997年第4期68-70,共3页 Marine Sciences
基金 山东省自然科学基金!Q95 B0 314号 国家自然科学基金资助项目!49376 2 78号
关键词 镉-铜法 硝酸盐 测定 过度还原 海水 Cd Cu method, Nitrate analysis, Excessive redution
  • 相关文献

参考文献2

  • 1团体著者,海洋调查规范,1991年,273页
  • 2韩舞鹰,海水化学要素调查手册,1986年,136页

同被引文献27

  • 1邱进坤,张树刚,姚炜民,郜钧璋,蔡圣伟.QuAAtrO连续流动分析仪测定海水中营养盐[J].环境科学与技术,2011,34(S2):187-189. 被引量:8
  • 2陈水土,戴民汉,苏国成,王本旺,黄韬.海水中硝酸盐铜镉柱还原测定操作步骤的优化研究[J].台湾海峡,2004,23(4):429-435. 被引量:10
  • 3刘菁,蒋丹凤.烘箱法测定地表水中的总氮[J].污染防治技术,2005,18(5):50-52. 被引量:8
  • 4程丽巍,许海,陈铭达,刘兆普,陈巍,华静霞,霍珊珊.水体富营养化成因及其防治措施研究进展[J].环境保护科学,2007,33(1):18-21. 被引量:79
  • 5中国国家标准化管理委员会.中华人民共和国国家标准,GB/T5009.33食品中亚硝酸盐与硝酸盐的测定方法[S].北京:中国标准出版社,2003.
  • 6中国国家标准化管理委员会.中华人民共和国国家标准,GB/T15401-1994水果、蔬菜及其制品亚硝酸盐和硝酸盐含量的测定[S].北京:中国标准出版社,2004.
  • 7中国国家标准化管理委员会.中华人民共和国国家标准,GB18406.1-2001农产品安全质量无公害蔬菜安全要求[S].北京:中国标准出版社,2001.
  • 8中国国家标准化管理委员会.中华人民共和国国家标准,GB/T5009.1-2003食品卫生检验方法理化部分总则[S].北京:中国标准出版社,2004.
  • 9中国国家标准化管理委员会.中华人民共和国国家标准,GB/T6379-1986测试方法的精密度通过实验室间标准测试方法的重复性和再现性[S].北京:中国标准出版社,1986.
  • 10OLDAY F C, BARKEER A V, MAYNARD D N, et al. Aphysiological basis for different patterns of nitrate accumulation in two spinach caltivars[J]. J Amer Soc Hort Sci, 1976,101 (3) : 217-219.

引证文献4

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部