摘要
认为用镉 -铜法测定硝酸盐时还原率低的原因是过度还原 ,并介绍了判断过度还原存在与否的实验方法 ,研究了影响过度还原程度的因素及解决方法。
Usually it is considered that the nitrate reduction column should be reactivated if the reduction efficiency is less than 95 %,because the reduction column has low reduction ability, but this study indicated that the low reduction efficiency may also caused by excessive reduction. A simple experiment method to determine whether excessive reduction has occurred was introduced, and the factors which influenced the excessive reduction degree were discussed.
出处
《海洋科学》
CAS
CSCD
北大核心
1997年第4期68-70,共3页
Marine Sciences
基金
山东省自然科学基金!Q95 B0 314号
国家自然科学基金资助项目!49376 2 78号