摘要 目的探讨不同交联浓度的抗癌胚抗原(CEA)单克隆抗体对液相芯片荧光检测信号的影响,明确CEA检测中最适抗体交联浓度。方法分别以不同交联浓度的标记有生物素的抗-CEA单抗与Luminex公司生产的44号荧光微球交联,交联完成后在每个反应孔中加入交联有不同浓度抗体的荧光微球各5000个,每个浓度加8孔,在每个抗体交联浓度下对应的检测孔中分别加入藻红蛋白(PE)标记的亲和素,经过15min的反应后用Luminex-100荧光检测仪检测荧光强度。结果CEA抗体的最适交联浓度为100μg/ml。结论在液相芯片的蛋白检测中CEA抗体有其最适的交联浓度。 Objective To explore the effect of different concentrations of coupled CEA antibody on liquichip fluorescence detection signal,and ascertain the optimal coupled CEA antibody concentration during CEA detection. Methods Anti-CEA monoclonal antibodies labelled with biotin at different concentrations (25, 50, 100, 150, 200μg/ml) were coupled to No. 44 fluorescent beads produced by Luminex Company. Each of 8 wells was added 5 000 beads coupled with the same concentration of antibody respectively. Different concentrations of avidin labeled with phycoerythrin (PE) were added the corresponding wells (0. 08, 0. 16, 0.32, 0.64, 2.56, 12.80, 64.00, 128.00 μg/ml corresponding to 25, 50, 100, 150,200 μg/ml of antibodies respectively). After incubation for 15 min under 37℃ ,Luminex-100 was used to measure fluorescence intensity. Results The optimal coupled anti-CEA antibody concentration was 100 μg/ml. Conclusion There exists the optimal coupled CEA antibody concentration during liquiehip protein detection.
出处 《国际检验医学杂志》 CAS 2007年第11期994-996,共3页 International Journal of Laboratory Medicine
关键词 CEA 抗体 液相芯片 交联效率 信号检测 CEA,antibody Liquichip Coupled efficiency Signal detection
  • 相关文献

参考文献5

  • 1孙凯,汪谦.液相芯片技术研究应用进展[J].中华实验外科杂志,2005,22(5):639-640. 被引量:13
  • 2Jia XC, Raya R, Zhang L, et al. A novel method of Multiplexed Competitive Antibody Binning for the characterization of monoclonal antibodies [J]. J Immunol Methods, 2004, 288(1-2): 91- 98.
  • 3Picketing JW, Martins TB, Greer RW, et al. A multiplexed fluorescent microsphere immunoassay for antibodies to pneumococcal capsular polysaccharides[J]. Am J Clin Pathol, 2002, 117 (4) : 589-596.
  • 4Wong SJ, Demarest VL, Boyle RH, et al. Detection of human anti-flavivirus antibodies with a west nile virus recombinant antigen microsphere immunoassay [J]. J Clin Microbiol, 2004, 42 (1): 65-72.
  • 5Bellisario R, Colinas RJ, Pass KA. Simultaneous measurement of thyroxine and thyrotropin from newborn dried blood-spot specimens using a multiplexed fluorescent microsphere immunoassay [J]. Clin Chem, 2000, 46(9):1422-1424.

二级参考文献16

  • 1Martins TB,Pasi BM,Litwin CM,et al.Heterophile antibody interference in a multiplexed fluorescent microsphere immunoassay for quantitation of cytokines in human serum.Clin Diagn Lab Immunol,2004,11:325-329.
  • 2Wong SJ,Demarest VL,Boyle RH,et al.Detection of human anti-flavivirus antibodies with a west nile virus recombinant antigen microsphere immunoassay.J Clin Microbiol,2004,42:65-72.
  • 3Jager W,Velthuis H,Prakken BJ,et al.Simultaneous detection of 15 human cytokines in a single sample of stimulated peripheral blood mononuclear cells.Clin Diagn Lab Immunol,2003,10:133-139.
  • 4Biagini RE,Sammons DL,Smith JP,et al.Comparison of a multiplexed fluorescent covalent microsphere immunoassay and an enzyme-linked immunosorbent assay for measurement of human immunoglobulin G antibodies to anthrax toxins.Clin Diagn Lab Immunol,2004,11:50-55.
  • 5Spiro A,Lowe M,Brown D.A bead-based method for multiplexed identification and quantitation of DNA sequences using flow cytometry.Appl Envir Microbiol,2000,66:4258-4265.
  • 6Bortolin SM,Black H,Modi I,et al.Analytical validation of the tag-it high-throughput microsphere-based universal array genotyping platform:application to the multiplex detection of a panel of thrombophilia-associated single-nucleotide polymorphisms.Clin Chem,2004,50:2028-2036.
  • 7Buranda T,Lopez GP,Simons P,et al.Detection of epitope-tagged proteins in flow cytometry:fluorescence resonance energy transfer-based assays on beads with femtomole resolution.Anal Biochem,2001,298:151-162.
  • 8Jani IV,Janossy G,Brown DW,et al.Multiplexed immunoassays by flow cytometry for diagnosis and surveillance of infectious diseases in resource-poor settings.Lancet Infect Dis,2002,2:243-250.
  • 9Prabhakar UE,Eirikis M,Reddy E,et al.Validation and comparative analysis of a multiplexed assay for the simultaneous quantitative measurement of Th1/Th2 cytokines in human serum and human peripheral blood mononuclear cell culture supernatants.J Immunol Methods,2004,291:27-38.
  • 10Faucher S,Martel A,Sherring A,et al.Protein bead array for the detection of HIV-1 antibodies from fresh plasma and dried-blood-spot specimens.Clin Chem,2004,50:1250-1253.

共引文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部