用锥形束CT图像测量放疗摆位误差 被引量:60

Setup errors with image guided radiotherapy technique
原文传递
导出
摘要 目的探讨图像引导放疗(IGRT)系统的临床摆位误差。方法应用医科达SynergyTM系统治疗51例,其中头颈部肿瘤19例,胸部肿瘤25例,腹、盆部肿瘤7例。每次照射前获取X线容积图像(XVI)即锥形束断层扫描(CBCT),将CBCT图像和计划CT图像及其靶中心匹配,获得靶中心x、y、z方向的误差,分析误差及其分布规律。结果CBCT图像对比度好,临床上能清楚显示肿瘤大小、位置及其和周围的关系。51例每次治疗开始前的CBCT共955次,分别分析头颈部、胸部及腹、盆部摆位误差,均呈正态分布。系统误差(均数)±随机误差(标准差)在x、y、z方向分别为头颈部(- 0.3±1.5)、(-0.3±1.6)、(0±1.3)mm,胸部(-0.6±2.9)、(-0.2±5.6)、(0.7±3.2)mm,腹、盆部(-0.8±2.1)、(-0.3±5.9)、(0.1±2.6)mm。结论通过CBCT测量治疗前摆位误差并实时调整摆位误差,可大大提高患者摆位精度,同时为正确设定计划靶体积(PTV)提供了依据。 Objective This work was to investigate the setup errors of image-guided-radiotherapy (IGRT) . Methods From April 2006 to August 2006, 51 patients were treated with the newly installed Elekta SynergyTM system, among whom there were 19 head and neck tumors, 25 thoracic tumors and 7 abdomino-pelvic tumors. All patients had received kilovoltage cone beam computed tomography (kv-CBCT) before radiotherapy was delivered. The acquired X-ray volumetric images (XVI) were co-registered with planning CT and errors of isocenter positions on x, y, z axes were obtained and analyzed. Results The acquired CBCT images had high quality clearly showing soft tissue organs, tumor size , location and the surrounding structures. All 51 patients received CBCT scan at each treatment fraction before radiation delivery. A total of 955 CBCT scans were obtained. Setup errors in the head and neck, chest and abdominopelvis all gave normal distributions. The systemic ± random errors on x, y, z directions were ( - 0.3 ± 1.5 ) mm, ( -0.3 ±1.6) mm,(0 ±1.3) mm for the head and neck;( -0.6 ±2.9) mm,( -0.2 ±5.6) mm,(0.7 ±3.2) mm for the chest; ( -0.8 ±2.1) mm,( -0.3 ±5.9) mm,(0.1 ±2.6) mm for the abdominopelvis, respectively. Conclusion Measurement of setup error before radiation delivery using kV-CBCT scan combined with on-line correction could greatly improve the patient setup precision and may guide optimal designing.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2007年第6期461-464,共4页 Chinese Journal of Radiation Oncology
基金 四川省科技攻关计划资助项目(03SG022-008 04SG022-007)
关键词 图像引导放疗 断层摄影术X线 计算机 锥形束 摆位误差 Image-guided radiotherapy Tomagraph,computer Conebeam Setup errors
  • 相关文献

参考文献15

  • 1Hurkmans CW, Remeijer P, Lebesque JV, et al. Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol, 2001,58 : 105-120
  • 2van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol,2004,14:52-64
  • 3Kutcher GJ, Mageras GS, Liebel SA. Control, correction, and modeling of setup errors and organ motion. Semin Radiat Oncol,1995,5 : 134-145
  • 4van Herk M, Remeijer P, Lebesque JV. Inclusion of geometric uncertainties in treatment plan evaluation. Int J Radiat Oncol Biol Phys, 2002,52 : 1407-1422
  • 5Matthias G, Juergen M, Dirk V, et al. Magnitude and clinical relevance relevance of translational and rotational patient setup errors : a cone-beam CT study. Int J Radiat Oncol Biol Phys,2006,5:934-942
  • 6Catherine AM, Ann MH, Jonathan S, et al. X-ray volumetric imaging in image-guided radiotherapy: the new standard in on-treatment imaging. Int J Radiat Oncol Biol Phys, 2006,64:625-634.
  • 7Michalski J, Purdy JA. Image-guided radiation therapy committee.Int J Radiat Oncol Biol Phys,2001,51:60-65
  • 8Mackie TR, Kapatoes J, Ruchala K, et al. Image guidance for precise conformal radiotherapy. Int J Radiat Oncol Biol Phys, 2003,56 : 89-105
  • 9Sidhu K, Ford EC, Spirou S, et al. Optimization of conformal thoracic radiotherapy using cone-beam CT imaging for treatment verification. Int J Radiat Oncol Biol Phys, 2003,55:757-767
  • 10Brock KK, Dawson LA, Sharpe MB, et al. Feasibility of a novel deformable image registration technique to facilitate classification, targeting, and monitoring of tumor and normal tissue. Int J Radiat Oncol Biol Phys, 2006,64:1245-1254

同被引文献578

引证文献60

二级引证文献296

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部