期刊文献+

高分辨连续光源原子吸收光谱仪检测饮用矿泉水中的金属离子 被引量:9

High-resolution Continuum Source Atomic Absorption Spectrometer and Its Application in Determination of Metal Ions in Drinking Mineral Water
下载PDF
导出
摘要 简要介绍了高分辨连续光源原子吸收光谱仪(HS—CS—AAS)的基本特点,并利用HS—CS—AAS测定了饮用矿泉水中的钙、镁、锶、钾、钠等微量元素。结果表明,测定各种离子的加标回收率为96.2%~100.9%,相对标准偏差(RSD,n=7)为0.44%~4.90%。 The basic characteristics of high-resolution continuum source atomic absorption spectrometer (HS-CS-AAS) is introduced in this paper and a method based on the instrument for the determination of trace metal elements of calcium, magnesium, strontium, potassium and sodium in drinking mineral water has been developed. The recovery of the method is 96.2% - 100.9% with precision of 0.44% - 4.90% RSD (n = 7).
作者 汪雨
出处 《岩矿测试》 CAS CSCD 2007年第6期485-489,共5页 Rock and Mineral Analysis
关键词 连续光源原子吸收光谱仪 高分辨 饮用矿泉水 微量元素检测 continuum source atomic absorption spectrometer high-resolution drinking mineral water trace element detection
  • 相关文献

参考文献13

  • 1烟伟.火焰原子吸收光谱法测定温泉矿泉水中微量元素[J].理化检验(化学分册),1999,35(11):513-514. 被引量:3
  • 2卢爱民,柴辛娜,高宏宇,胡圣虹,帅琴.溶剂萃取-石墨炉原子吸收光谱法测定水样中的痕量铅[J].分析科学学报,2006,22(2):190-192. 被引量:19
  • 3黄晖,肖珊美,陈建荣.PAN浊点萃取-火焰原子吸收光谱法测定水样中的痕量钴[J].光谱实验室,2005,22(5):1003-1006. 被引量:21
  • 4James M Harnly. The Future of Atomic Absorption Spectrometry: A Continuum Source with a Charge Coupled Array Detector[ J ]. J Anal At Spectrom, 1999,14 : 137 - 146.
  • 5Becker-Ross H, Florek S, Heitmann Uwe. Observation, Identification and Correction of Structured Molecular Background by Means of Continuum Source AAS-- Determination of Selenium and Arsenic in Human Urine [J]. J Anal At Spectrom ,2000,15 : 137 - 141.
  • 6Borges Daniel L G, da Silva A F, Welz Bernhard, et al. Determination of Lead in Biological Samples by High-resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry with Direct Solid Sampling[ J ]. J Anal At Spectrom ,2006,21:763 - 769.
  • 7Lepri F G, Welz Bernhard, Borges D L G, et al. Speciation Analysis of Volatile and Non-volatile Vanadium Compounds in Brazilian Crude Oil Using High-resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry [ J ]. Analytica Chimica Acta, 2006,558 : 195 - 200.
  • 8da Silva A F, Lepri F G, Borges D L G, et al. Determination of Mercury in Biological Samples Using Solid Sampling High-resolution Continuum Source Electrothermal Atomization Atomic Absorption Spectrometry with Calibration Against Aqueous Standards [ J ]. J Anal At Spectrom , 2006 , 21:13 21 - 1326.
  • 9Huang M D, Becker-Ross H, Florek S, et al. Determination of Phosphorus by Molecular Absoption of Phosphorus Monoxide Using a High-resolution Continuum Source Absorption Spectrometer and an Air-Acetylene Flame[ J ]. J Anal At Spectrom,2006 ,21 : 338 - 345.
  • 10Huang Mao Dong, Becker-Rossa H, Florek S, et al. Determination of Sulfur by Molecular Absorption of Carbon Monosulfide Using a High-resolution Continuum Source Absorption Spectrometer and an Air-Acetylene Flame [ J ]. Spectrochimica Acta Part B: Atomic Spectroscopy,2006,61 (2) : 181 - 188.

二级参考文献12

共引文献39

同被引文献274

引证文献9

二级引证文献113

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部