摘要 讨论了一种在核磁共振成像数字谱仪中保证射频发射机与接收机之间射频信号相位相干的方法,该方法在成像脉冲序列执行过程中,同步切换发射机与接收机的频率,并在脉冲序列最后,将发射和接收频率恢复到脉冲序列最初的频率值,从而实现发射与接收的信号相干,其优点是无需加入额外的"回绕延时",从而使得脉冲序列的编写得到简化,并且该方法不依赖于磁共振谱仪发射机和接收机硬件部分的具体结构,是适用于数字化谱仪的一种普遍的方法. Phase coherence between RF transmitter and receiver in digital MRI spectrometer was studied. In order to keep phase coherence between the transmitter and the receiver, we implemented a method in which the frequencies of the transmitter and the receiver were switched simultaneously during pulse sequence execution, and reset to their original values, respectively, at the end of sequence. The experimental results showed that this method is suitable for digital spectrometer and independent on the structure of transmitter and receiver.
出处 《波谱学杂志》 CAS CSCD 北大核心 2007年第4期439-445,共7页 Chinese Journal of Magnetic Resonance
基金 上海市科委科研计划项目资助(06DZ22014 06DZ11315)
关键词 磁共振成像 射频发射 射频接收 相位相干 magnetic resonance imaging, RF transmitter, RF receiver, phase coherence
  • 相关文献

参考文献5

  • 1DRX programming user manual[Z].resonance instruments Ltd, 2002.
  • 2Michal Carl A, Broughton Kesten, Hansen Elsa. A high performance digital receiver for home-built nuclear magnetic resonance spectrometers [J]. Rev Sci Instrum, 2002, 73: 453-458.
  • 3徐勤,沈杰,蒋瑜,李鲠颖.一体化核磁共振数字接收机的设计[J].波谱学杂志,2005,22(4):357-365. 被引量:12
  • 4宁瑞鹏,徐勤,沈杰,李鲠颖.基于PCI总线的全功能数字化频率源[J].波谱学杂志,2006,23(2):209-215. 被引量:7
  • 5Zu Dong-lin(俎栋林).Magnetic resonance imaging(核磁共振成像学)[M].Bingjing(北京):Higher Education press(高等教育出版社),2004.

二级参考文献20

  • 1Wittebort R J, Woehler S E, Bradley C H. Efficient probe and receiver for solid-state 2H NMR[J]. J Magn Reson, 1986,76:143-148.
  • 2Villa M, Tian F, Cofrancesco P, et al. High-resolution digital quadrature detection[J]. Rev Sci Instrum, 1996, 67:2123-2129.
  • 3Li G Y, Xie H B. Digital quadrature detection in nuclear magnetic resonance spectroscopy[J]. Rev Sci Instrum, 1999, 70:1511-1513.
  • 4Michal C A, Broughton K, Hansen E. A high performance digital receiver for home-built NMR spectrometers[J]. Rev Sci Instrum, 2002, 73:453-458.
  • 5Moskau D. Application of real time digital filters in NMR spectroscopy[J]. Concepts in M R , 2002, 15:164-176.
  • 6Hoenninger J C, Crooks L E, Jones J W. Synchronized digital signal processor for MRI reception[P]. US:5529068 1996-06-25.
  • 7Li Y G, Jiang Y, Yan X L, et al. Digital nuclear magnetic resonance spectrometer[J]. Rev Sci Instrum, 2001, 72:4460-4463.
  • 8Wright S M, Brown D G, Proter J R, et al. A desktop magnetic resonance imaging system[J]. Magn Reson Mat Phys Biol Med(MAGMA), 2002, 13:177-185.
  • 9Momo F, Sotgiu A, Testa L, et al. Digital frequency synthesizers for nuclear magnetic resonance spectroscopy[J]. Rev SciInstrum, 1994, 65(10):3291-3292.
  • 10DXR Programming User Manual. Resonance Instruments Inc[M]. Oxfordshire:United Kingdom, 2002.