介入超声印压系统检测超声仿体硬度的实验研究 被引量:2

Experimental study on assessment of ultrasound phantoms stiffness by intervention ultrasound indention system
原文传递
导出
摘要 目的评价介入超声印压系统检测不同硬度超声仿体的可行性和准确性。方法制作3组不同硬度的超声仿体,单轴压缩检测其杨氏模量作为标准,同时应用介入超声印压系统检测其仿体的硬度。结果两种方法检测的第1组仿体硬度值分别为(21.8±5.1)kPa,(45.5±5.7)kPa;第2组仿体硬度值分别为(44.8±7.1)kPa,(86.1±15.0)kPa;第3组仿体硬度值分别为(85.2±7.7)kPa,(117.9±10.9)kPa。两种方法检测硬度值的相关系数r=0.94。结论介入超声印压系统能够检测不同硬度的超声仿体,其硬度值与杨氏模量具有强相关,该仪器具有检测活体心肌和深部组织硬度变化的潜力。 Objective To explore the feasibility and accurate of interventional ultrasound indention system detecting phantoms stiffness.Methods The three different stiffness phantoms were made and their Young's module were measured by uniaxial ultrasound compression system.The ratio of the indentation pressure to the phantoms relative deformation was obtained from intervention ultrasound indention.ResultsThese phantoms Young's module measured by uniaxial ultrasound compression system were(21.8±5.1)kPa,(44.8±7.1)kPa,(85.2±7.7)kPa,respectively and their stiffness were(45.5±5.7)kPa,(86.1±15.0)kPa,(117.9±10.9)kPa,respectively by intervention utrasound indention system.The stiffness was well correlated with the Young's modulus(r=0.94).Conclusions The intervention ultrasound indention system may detect different stiffness phantoms and it may be also applied for the stiffness assessment of myocardium and deep tissue in vivo.
出处 《中华超声影像学杂志》 CSCD 2007年第11期1000-1002,共3页 Chinese Journal of Ultrasonography
关键词 超声检查 介入性 仿生材料 硬度 Ultrasonography,interventional Biomimetic materials Hardness
  • 相关文献

参考文献10

  • 1Madsen EL, Hobson MA, Shi H, et al. Tissue-mimicking agar/ gelatin aterials for use in heterogeneous elastography phantoms. Phys Med Biol, 2005,50 : 5597-5618.
  • 2Madsen EL, Frank GR, Hobson MA, et al. Spherical lesion phantoms for testing the performance of elastography systems. Phys Med Biol,2005,50:5983-5995.
  • 3Zheng YP,Choi YK,Wong K, et al. Biomechanical assessment of plantar foot tissue in diabetic patients using an ultrasound indentation system. Ultrasound Med Biol,2000,26:451-456.
  • 4Lu MH, Zheng YP, Huang QH. A novel noncontact ultrasound indentation system for measurement of tissue material properties using water jet compression. Ultrasound Med Biol, 2005,31 : 817- 826.
  • 5Mandinov L, Eberli FR, Seiler C, et al. Diastolic heart failure. Cardiovasc Res, 2000,45 : 813-825.
  • 6Zile MR, Brutsaert DL. New concepts in diastolic dyfunction and diastolic heart failure. Part Ⅰ: diagnosis, prognosis, and measurements of diastolicfunction. Circulation, 2002, 105:1387-1393.
  • 7Leung SF, Zheng YP,Choi CY, et al. Quantitative measurement of post irradiation neck fibrosis based on the young modulus:description of a new method and clinical results. Cancer, 2002,95 : 656-662.
  • 8Kawchuk GN, Fauvel OR. Sources of variation in spinal indentation testing:indentation site relocation,intraabdominal pressure, subject movement, muscular response, and stiffness estimation. J Manip Physiol Ther,2001,24 : 84-91.
  • 9Suh JKF,Youn I, Fu FH. An in situ calibration of an ultrasound transducer:A potential application for an ultrasonic indentation test of articular cartilage. J Biomech, 2001,34 :1347-1353.
  • 10Laasanen MS, Toyras J, Hirvonen J, et al. Novel mechano-acoustic technique and instrument for diagnosis of cartilage degeneration. Physiol Meas, 2002,23 : 491-503.

同被引文献29

  • 1邓辉胜,黄晶,邓昌明,刘地川,江永红,李进嵩,袁侨英,朱悫,黄如斌,蒋伟,王志刚,蔡恒辉,郭晋峰,王兴雄,杨金耀,郑永平.活体心肌力学属性的介入超声导管检测:仪器设计[J].中国医学影像技术,2006,22(10):1543-1546. 被引量:4
  • 2邓辉胜,黄晶,郑永平,邓昌明,刘地川,李进嵩,王志刚.介入超声印压检测活体心肌硬度的实验研究[J].中国超声医学杂志,2007,23(3):164-166. 被引量:6
  • 3Lin D, Ilkhanoff L, Gerstenfeld E, et al. Twelve-lead electrocardiographic characteristics of the aortic cusp region guided by intracardiac echocardiography and electroanatomic mapping. Heart Rhythm, 2008,5(5) :663-669.
  • 4Alaeddini J, Wood MA, Parvez B, et al. Site localization and characterization of pain during radiofrequency ablation of the pulmonary veins. Pacing Clin Electrophysiol, 2007,30(10) :1210-1214.
  • 5Rigatelli G, Dell'Avvocata F, Roncon L, et al. Intracardiac echocardiography guided atrial septal defect transcatheter closure of a very old surgical residual shunt. Minerva Cardioangiol, 2008, 56 (1) :171-175.
  • 6Gentry KL, Smith SW. Integrated catheter for 3-D intracardiac echocardiography and ultrasound ablation. IEEE Trans Ultrason Ferroelectr Freq Control, 2004,51(7) :800-808.
  • 7Stephens DN, Cannata J, Liu R, et al. The acoustic lens design and in vivo use of a multifunctional catheter combining intracardiac ultrasound imaging and electrophysiology sensing. IEEE Trans Ultrason Ferroelectr Freq Control, 2008,55(3):602-618.
  • 8Ramos M, Depasquale E, Coplan NL. Assessment of myocardial viability: review of the clinical significance. Rev Cardiovasc Med, 2008,9(4) :225-231.
  • 9Cooper JM, Epstein LM. Use of intracardiac echocardiography to guide ablation of atrial fibrillation [J]. Circulation, 2001, 104(25): 3010-3013.
  • 10Epstein LM, Mitchell MA, Smith TW, et al. Comparative study of fluoroscopy and intracardiac echocardiographic guidance for the creation of linear atrial lesions [J]. Circulation, 1998, 98 (17): 1796-1801.

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部