摘要 基于化学发光及氧对荧光的猝灭作用,开发了一种荧光氧传感器(传感系统)。考虑到气体猝灭剂(氧)在高分子半透膜内同时存在吸附与溶解作用,一个新的荧光猝灭动力学模型建立起来了。该方程能够很好地说明荧光猝灭的机理,并能够非常准确地拟合实验数据。 A novel optical oxygen sensor (sensing system) based on the chemical luminescene and oxygen quenching of luminescence has been developed. A new dynamic quenching model which combined the qurnching kinetics and the solubility and adsorption equations of quencher (oxygen) in polymer is proposed to correlate the decreases of luminescence intensity with the quencher concentration. This model can explain the common features of luminescent quenchung reactions in a polymer matrix and fit experimental data equally well.
出处 《传感技术学报》 CAS CSCD 1997年第3期22-26,共5页 Chinese Journal of Sensors and Actuators
基金 国家自然科学基金 国家教委留学回国人员基金 广东省自然科学基金资助项目
关键词 荧光传感器 发光 猝灭 浓度 检测 optical sensor luminescence quenching oxygen dynamics
  • 相关文献

参考文献2

  • 1阮复昌,华南理工大学学报,1994年,22卷,3期,32页
  • 2李向明,Transient DO Measurement Using a Computerized Membrance Electrode,1987年,213页

同被引文献22

  • 1Karube I, Matsunaga T, Mitsuda S, et al. Microbial electrode BOD sensor[J]. Biotechnol Bioeng, 1977, 19(10) : 1535 - 1547.
  • 2Lin Ling, Xiao Lai-long, Huang Sha, et al. Novel BOD optical fiber sensor based on co-immobilized microorganisms in ormosils matrix [J ]. Biosensors and Bioelectronics, 2006, 21 (9) : 1703 - 1709.
  • 3Xin Ling-ling, Wang Xu-dong, Guo Guang-mei, et al. An optical biosensing film for biochemical oxygen demand determination in seawater with an automatic flow sampling system [J]. Measuremnt Science and Technology, 2007, 18 : 2878 - 2884.
  • 4Yang Z, Suzuki H, Sasaki S, et al. Disposable sensor for biochemical oxygen demand [ J ]. Appl Microbiol Biotechnol, 1996, 46 : 10 - 14.
  • 5Li You-Long, Chu Ju. Study of BOD microbial sensors for the waste water treatment control [J]. Applied Biochemistry and Biotechnology, 1991, 28 : 855 - 863.
  • 6Tammeveski K, Kikas T, Tenno T, et al. Preparation and characterization of phtinum coatings for long life-time BOD biosensor [J]. Sensors and Actuators B, 1998, 47:21 -29.
  • 7Chee Gab-Joo, Nomura Yoko, Karube Isao. Bioscnsor for the estimation of low biochemical oxygen demand[J]. Analytica Chimica Acta, 1999, 379:185 - 191.
  • 8Liu Jing, Bjornsson Lovisa, Mattiasson Bo. Immobilised activated sludge based biosensor for biochemical oxygen demand measurement [ J ]. Biosensors and Bioelectronics, 2000, 14: 883 - 893.
  • 9Jung Jaechoon, Lee Byoungho. Erbium-doped fiber temperature sensor with high resolution using 1480nm pump laser. Proc. SPIE, 1999, 3746:538-541.
  • 10Wade S A, Muscat J C, Collins S F et al.. Nd^3+ doped optical fiber lemperature sensor using the fluorescence intensity ratio lechnique. Rev. Sci. Instruments, 1999, 70(11) :4279-4282.