摘要
目的介绍一种新的颈椎前路植骨打骨器。方法对传统的颈椎前路植骨打骨器的头部做了一个厚2㎜,直径10㎜的圆形台阶。结果颈椎植骨块的位置能被下陷1~2㎜,没有发现植骨块脱落的情况。结论这种新的颈椎前路植骨打骨器使用相对简单、安全、可靠。适用于颈椎前路植骨块的打入。
Objective To introduce a new impacted bone apparatus for anterior cervical spine graft bone. Methods Top of traditional impacted bone apparatus for anterior cervical spine graft bone was made a circle step with 2-mm-thick. Its diameter is 10 mm Results Position the graft bone in the cervical spine was sink 1 ~ 2 mm. No the graft bone was take off. Conclusion The new impacted bone apparatus is relatively simple, reliable and safe to use and suitable for anterior cervical spine graft bone impaction.
出处
《生物骨科材料与临床研究》
CAS
2007年第6期52-52,共1页
Orthopaedic Biomechanics Materials and Clinical Study
关键词
器械
颈椎
植骨
Apparatus
Cervical spinet Graft bone