摘要
为提高诊断X线机的诊断质量,促进放射防护工作的开展,根据"国际辐射单位和测量委员会"(International Commission on Radiological Units and Measurements,ICRU),"国际放射防护委员会"(International Commission on Radiation Protection,ICRP)关于辐射对人体的损害之划分标准,作者对岛津制作所90年代中期生产的XEB150L-20型500mA X线机按照国家规定的要求及有关标准,分别对辐射空气释动能率,半价层,输入量重复性,输出量线性,高对比分辨率,光野与照射野一致性及X线管的焦点等参数进行了测试。分析了测试的数据与结果,并进行了简要的评价。
In order to improve the diagnostic quality of X-ray device and to promote the implementation of radiation protection in accordance to the ICRU definition and ICRP classification standards for measuring radiation harm to human body,we have made a number of tests, including the kinetic energy released by ionizing air, the Half-value layer, the output dose coincidence, the High contrast resolution, the Consistency between visual light field and project X-ray field, and the Focus of X-ray tube. We have also analysed the measured data and results. The tests were conducted on the XEB150L - 20 made by Shimadzu Co. in the middle of 1990s.