SIEMENS MD7745加速器射野大小对电子束输出因子的影响

The influence of different radiated fields from SIEMENS MD7745 accelerator on output factors of electrons beam
下载PDF
导出
摘要 目的研究射野大小对高能电子束输出因子的影响。方法用Capintec292剂量测量系统测量SIMENS MD7745直线加速器9Mev和12 Mev电子辐射源各种射野成形情况下的输出因子,并进行数学分析,为临床准确使用该因子提供依据。结果得到了各种照射情况下的电子束输出因子,可知该因子对准直铅门大小、限光筒大小、铅档野开口大小和能量有很强的依赖性。结论为了保证剂量计算精度达到WHO规定的范围,临床实际工作中,限光筒大小应尽量接近实际所需射野,当射野边长大于电子束射程时,可直接使用标准限光筒输出因子,而对边长小于电子束射程的射野,必须有针对性地实际测量所用射野的输出因子。 Objective To study the influence of different radiated fields from accelerator on output factors of high-energy electron beam. Methods The influence of radiated fields from accelerator SIEMENS MD7745 of 12 Mev and 9Mev on output factors of high energy electron beam was measured by using 292 dosimetry systern. The data was analyzed to deduce the proper rules for clinical practice. Results The output factors gotten under different conditions were found to be strongly related to the collimator jaws,standard applicator sizes,lead block fields and electron energies. Conclusion In order to attain the rules required from WHO for quality assurance in radiotherapy, the applicator size must be close to the lead block field, on this basis the output factors can be used for standard applicators to replace the lead block field when the field size is larger than Rp, but the output factors must be measured for those fields which its sizea are less than Rp.
出处 《同济大学学报(医学版)》 CAS 2007年第6期22-24,33,共4页 Journal of Tongji University(Medical Science)
关键词 直线加速器 高能电子束 剂量测量 输出因子 linear accelerator high-energy electron dosimetry output factor
  • 相关文献

参考文献5

二级参考文献12

  • 1张红志 见:胡逸民 主编.高能电子束射野剂量学[A].见:胡逸民,主编.肿瘤放射治疗学[C].北京:原子能出版社,1999.253-254.
  • 2PALTA J R,DAFTARI I K,AYYANGAR K M,et al. Electron beam characteristics on a philips SL75[J].Med Phys,1990,17:23-24.
  • 3M.D.Mills,K.R.Hostron, P.R.Almond. Prediction of electron beam output factors[J]. Med. Phys, 1982,9.
  • 4Jatinder R.Palta,Iwder K .Daftari,K.M.Ayyanger,et al. Electron beam characteristics on a Philips SL2[J]. Med. Phys, 1990,17(1):27-34.
  • 5Inalsingh CH. An experience in treating 501 patients with keloids Johns Hopkins. Med J, 1974,134:284-290.
  • 6胡逸民.高能电子束剂量计算模型[A]..肿瘤放射物理学[C].北京:原子能出版社,1999.506-515.
  • 7Khan FM. The physics of radiation therapy. 2nd ed. Baltimore:Williams & Wilkins, 1994.369-371.
  • 8Almon SS, Samuel ST, Carl EN, et al. Comprehensive analysis of electron beam central axis dose for a radiotherapy linear accelerator. Med Phys, 1994,21: 559-566.
  • 9黄晓延,钱剑扬,钟宁山,卢杰,黄劭敏.高能电子线射野中心轴百分深度剂量测定[J].癌症,2000,19(6):603-606. 被引量:6
  • 10何京学,胡斌,张清文,朱庙生.电子束照射野面积对中心轴剂量和输出因子的影响[J].中华放射肿瘤学杂志,2001,10(3):197-199. 被引量:10

共引文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部