动态轮廓眼压计与非接触眼压计的比较 被引量:1

Comparison of dynamic contour and non-contact tonometer
下载PDF
导出
摘要 目的比较动态轮廓眼压计(DCT)与非接触眼压计(NCT)测量眼压的差异,并探讨中央角膜厚度(CCT)对这两种测量方法的影响。方法对75例拟接受近视手术的患者进行角膜厚度测量后,对所有眼分别用DCT和NCT进行眼内压(IOP)的测量,并比较中央角膜厚度与两种眼压计读数的关系。对所有数据进行t检验和相关分析。结果NCT和DCT测得的眼压值具有显著性差异(t=9.2932,P<0.01)。NCT测量值与CCT呈正相关,相关系数r=0.3482(P<0.01)。DCT测量值与CCT无相关性,相关系数r=0.0635,P>0.05)。结论NCT和DCT测量眼压值有差异,DCT测量值大于NCT,CCT对NCT的影响大于DCT。 Objective To compare the difference of intraocular pressure measured by non - contact tonometer (NCT) and Dynamic contour tonometer(DCT) and to research the influence of central corneal thickness on the measurements of IOP obtained with NCT and DCT. Method Central corneal thickness, IOP measured by NCT and DCT were measured on 75 (150 eyes) patients before refractive surgery. The two kinds of IOP readings and its relationship with CCT were compared and analyzed by t-test and correlation analysis. Results There is significant difference between IOP measured by NCT and IOP by DCT ( t = 9.2932, P 〈 0.01 ). CCT was positive relative to the IOP measured by NCT; CCT was not relative to the IOP measured by DCT. The correlated coefficients are r =0.3482 ( P 〈0.01 ) and r =0.0635 ( P 〉 0.05 ) respectively. Conclusion There is difference between IOP measured by NCT and IOP by DCT. IOP measured by DCT was higher than which measured by NCT. The influence of central corneal thickness on the measurements of intraocular pressure with NCT was bigger than DCT.
出处 《临床眼科杂志》 2007年第6期519-521,共3页 Journal of Clinical Ophthalmology
关键词 眼内压 中央角膜厚度 动态轮廓眼压计 非接触眼压计 Intraocular pressure Central corneal thickness Non-contact tonometer Dynamic contour tonometer
  • 相关文献

参考文献10

  • 1俞振飞,方春庭,吴亚明,金成鹏.凸面眼压计与Goldmann压平眼压计C值测量结果比较[J].临床眼科杂志,2007,15(5):401-403. 被引量:1
  • 2Brandt JD. Corneal thickness in glaucoma screening, diagnosis, and management. Curt Opin Ophthalmol ,2004,15:85-89.
  • 3Duba I, Wirthlin AC. Dynamic contour tonometry for poot-LASIK intraocular pressure measurements. Klin Monatsbl Augenheilkd, 2004, 221:347-350.
  • 4张龄洁,管涛,卢丽蓉,邱朝霞.LASIK术前、术后角膜厚度与眼压的关系及其意义[J].临床眼科杂志,2006,14(5):399-401. 被引量:5
  • 5鲍捷,代艳,陈晓明.角膜厚度对两种眼压测量方法的影响[J].中国实用眼科杂志,2005,23(3):272-274. 被引量:15
  • 6Bron AM, Creuzot-Garcher C, Goudeau-Boutillon S, et al Falsely elevated intraocular pressure due to increased central corneal thickness. Grade's Arch Clin Exp Ophthalmol , 1999, 237:220-224.
  • 7Gordon MO, Beiser JA, Brandt JD, et al. The ocular hypertension treatment study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol, 2002, 120:714-720.
  • 8Shah S, Chatterjee A, Mathai M, et al. Relationship between cornealthickness and measured intraocular pressure in a genera/ophthalmology clinic. Ophthalmology, 1999, 106:2154-2160.
  • 9Feltgen N, Leifert D, Funk J. Correlation between central corneal thickness, applanation tonometry, and direct intracameral lOP readings. Br J Ophthalmol , 2001,85:85-87.
  • 10钟一声,龚洁,叶纹,王康孙.非接触式眼压计与Goldm ann压平眼压计测量眼压的比较[J].眼视光学杂志,2000,2(2):101-103. 被引量:32

二级参考文献26

  • 1鲍捷,代艳,陈晓明.角膜厚度对两种眼压测量方法的影响[J].中国实用眼科杂志,2005,23(3):272-274. 被引量:15
  • 2晏晓明,李海丽,王捷.三种眼压计在准分子激光原位角膜磨镶术后的应用比较[J].中国实用眼科杂志,2005,23(6):612-615. 被引量:9
  • 3胡铮.气眼压计的临床应用[J].中华眼科杂志,1986,22:5-5.
  • 4胡茂生 刘淳诗.非接触式眼压计与Goldmann氏压平眼压计对比研究[J].中华眼科杂志,1985,:111-111.
  • 5Shieds MB, The non-contact tonometer, Its value and limitations,Surv Ophthalmol, 1980, 24:211.
  • 6Whitacre MM et al. Am J Opthalmol, 1993, 115:592.
  • 7Ehlers N, BransenT, Sperling S, Applanation tonometry and central corneal thickness Acta Ophthalmol, 1975, 53: 34- 43.
  • 8Wolfs RCW et al. Am J Ophthalmol, 1997, 123:767.
  • 9Shah S, Ophthalmolgy, 2000, 107:1805.
  • 10Johnson M et al. Arch Ophthalmol, 1978, 96:664.

共引文献42

同被引文献8

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部