二次电子倍增器的非线性效应研究

Research on nonlinearity effect of secondary electron multiplier
下载PDF
导出
摘要 应用UTB-500铀同位素标准物质、多接收技术研究了一台热电离质谱计二次电子倍增器(SEM)的非线性效应。得出SEM的非线性效应在整个离子计数范围内都存在,非线性在离子计数率约为20000cps时,存在一个极值点,极值点的测量值与标准值之间的偏差可达3%,在极值点两边的非线性服从对数线性规律。提出了一种非线性校正的数学模型,应用该模型,可以校正热电离质谱计SEM非线性效应。 The nonlinearity of secondary electron multiplier(SEM) of a thermal ionization mass spectrometer has been researched by using UTB-500 uranium isotope reference material and multi-collecting technique.The results show that the nonlinearity effect of SEM exists in the whole ion counting range,and there is an extreme point of the nonlinearity when the ion counting rate is about 20000cps.The deviation between measured value of the extreme point and the reference value of the reference sample can be up to 3%,and the nonlinearity obeys logarithm linearity law on both sides of extreme point.A kind of mathematics model of nonlinearity calibration has been put forward.Using this model,the nonlinearity of SEM of TIMS can be calibrated.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2007年第5期927-932,共6页 Nuclear Electronics & Detection Technology
关键词 二次电子倍增器 非线性 校正 secondary electron multiplier nonlinearity calibration
  • 相关文献

参考文献9

  • 1复旦大学.原子核物理实验方法[M].北京:原子能出版社,1996.
  • 2M. E. Wieser, J. B. Schwieters. The development of multiple collector mass spectrometry for isotope ratio measurements. Int. J. Mass Spectrom. [J], 2005, 242: 97.
  • 3刘君红,刘俊勇,管兴胤.光电倍增管线性特性、时间特性参数调试[J].核电子学与探测技术,2005,25(6):768-771. 被引量:12
  • 4丁先利,王金川1,祁辉荣,肖国青,徐瑚珊,郭忠言,詹文龙,武丽杰,徐治国,孙志宇,李加兴,胡正国.时间型和增益型光电倍增管分压系统的比较研究[J].核电子学与探测技术,2005,25(4):392-395. 被引量:2
  • 5孟祥承,杨长根,王志民,石峰,关梦云,钟玮丽,刘金昌.光电倍增管的光电子幅度谱测试研究[J].核电子学与探测技术,2005,25(6):594-600. 被引量:7
  • 6S. Richter, et al. Ltnearity tests for secondary electron multipliers used in isotope ratio mass spectrometry[J]. Int. J. Mass Spectrom, 1999, 206: 105.
  • 7D. L. Hoffmann, et al. Characterisation of secondary electron multiplier nonlinearity using MCICPMS[J]. Int. J. Mass Spectrom, 2005, 244:97.
  • 8H. Rameback, et al.On the determination of the true dead time of a pulse-counting system in isotope ratio mass spectrometry [J]. J. Anal. At. Spectrom, 2001, 16: 1271.
  • 9K. L. Ramakumar, et al.Calibration procedures for a multicollector mass spectrometer for cup efficiency, detector amplifier linearity, and isotope fractionation to evaluate the accuracy in the total evaporation method [ J]. Int. J. Mass Spectrom, 1999 184:109.

二级参考文献8

  • 1王金川,肖国青,郭忠言,詹文龙,徐治国,祁辉荣,陈志强,武丽杰,丁先利,徐瑚珊,孙志宇,李加兴,李琛,高辉,张力,王猛,胡正国,陈立新.RIBLL终端LASCAR闪烁体阵列探测器研制进展[J].原子核物理评论,2005,22(1):84-87. 被引量:3
  • 2Kuhlen M et al, IEEE Trans. Nucl. Sci. ,1991,38(5):102
  • 3滨松光子学株式会社.光电倍增管-基础及应用[M].,1993..
  • 4WANG Quan-jin, et al. Nuclear Techniques, 2004,Vol.27,NO.3:233
  • 5Katsushi Arisaka arisaka@physics, ucla. edu Principle of PMT and its Calibration.
  • 6Bellamy EH. Absolute calibration and monitoring of a spectrometric channel using a photomultiplier[J]. Nucl Instrum Methods,1994, A339:468.
  • 7Dossi R, et al. Methods for precise photoelectron counting with photomultipliers[J]. Nucl Instrum Methods, 2000, A451 : 623.
  • 8周莉.闪烁光纤量能器的预制研究[C]..第十一届全国核电子学与核探测技术年会论文集[A].,2002.240.

共引文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部