种植体-基台连接形式对种植体周围骨组织应力分布的影响 被引量:12

Influence of implant-abutment connection mode on stress distribution in peri-implant bone
原文传递
导出
摘要 目的分析种植体-基台连接形式对种植体周围骨组织应力分布的影响,从生物力学角度探讨平台转换连接形式防止或减少种植体周围骨吸收的可能机制。方法利用COSMOSM2.85软件包建立种植体支持的下颌第一磨牙三维有限元模型,种植体-基台的连接形式分别采用平齐对接(模型A)和平台转换(模型B)。采用垂直和斜向两种形式加载,载荷均为200N,比较两种模型种植体周围骨组织的应力分布情况以及种植体-骨界面颊舌侧相同位置的von Mises应力大小。结果不同加载条件下两种模型种植体周围骨组织应力集中在种植体颈部颊舌侧骨皮质内,斜向加载时最大von Mises应力值高于垂直加载时。模型A和模型B骨组织内最大von Mises应力值在垂直加载时,分别为11.61MPa和7.15MPa,斜向加载时分别为22.07MPa和11.87MPa。距离种植体-基台连接处越远,von Mises应力值越小,骨皮质到骨松质交界处的应力变化最明显。与模型A相比,模型B种植体-骨界面相同节点的最大von Mises应力值较小。结论与平齐对接形式相比,平台转换设计可改善种植体周围骨组织的应力分布,降低种植体颈部骨组织所受的应力。 Objective To investigate the influence of butt joint connection and platform switching design of implant-abutment connection on the stress distribution in peri-implant bone. Methods Three- dimensional finite element models of implant-supported mandibular first molar with different implantabutment connections were computed by COSMOSM 2. 85. Traditional butt joint connection was used in model A and platform switching design in model B. Loading conditions were a vertical load of 200 N and inclined load of the same magnitude at 45 degrees to the vertical axis of the implant. Stress distribution in peri-implant bone and von Mises stresses at the same point of buccal and lingual implant-bone interfaces in two models were compared. Results Stresses concentrated in peri-implant cortical bone at the neck of implants on the buccal and lingual sides ; maximum von Mises stresses under inclined load were higher than those under vertical load. Maximum von Mises stress in bone was 11.61 MPa in model A and 7. 15 MPa in model B under vertical load, and 22. 07 MPa in model A and 11.87 MPa in model B under inclined load respectively. Von Mises stresses decreased as the distance from implant-abutment junction increased and the most obvious change occurred at the interface between cortical bone and spongy bone. Von Mises stresses at the same points of buccal and lingual implant-bone interfaces in model A were higher than those in model B. Conclusions Compared with butt joint connection, platform switching design improved the stress distribution and decreased the maximum stresses in peri-implant bone around implant cervix.
出处 《中华口腔医学杂志》 CAS CSCD 北大核心 2008年第1期50-53,共4页 Chinese Journal of Stomatology
关键词 牙种植 应力 物理 有限元分析 种植体-基台连接形式 Dental implantation Stress, mechanical Finite element analysis Implantabutment connection
  • 相关文献

参考文献12

  • 1Lazzara RJ, Porter SS. Platform switching: a new concept in implant dentistry for controlling post restorative crestal bone levels. Int J Periodontics Restorative Dent ,2006,26 (1) :9-17.
  • 2Baumgarten H, Coechetto R, Testori T, et al. A new implant design for erestal bone preservation: initial observations and ease report. Pract Proeed Aesthet Dent ,2005,17 (10) :735-740.
  • 3Gardner DM. Platform switching as a means to achieving implant esthetics. N Y State Dent J,2005,71 (3) :34-37.
  • 4谭荣,施斌.封闭介质对粘结固定型种植义齿应力分布的影响[J].口腔医学研究,2004,20(1):22-25. 被引量:4
  • 5Chun HI, Shin HS, Han CH, et al. Influence of implant abutment type on stress distribution in bone under various loading conditions using finite element analysis. Int J Oral Maxillofac Implants, 2006,21 (2) : 195-202.
  • 6李智勇,刘学军,赵丽雅,胡建,程祥荣.不同基台时种植体支持全瓷单冠的应力分析[J].口腔医学研究,2007,23(4):402-404. 被引量:16
  • 7Kitamura E, Stegaroiu R, Nomura S, et al. Biomechanical aspects of marginal bone resorption around osseointegrated implants: considerations based on a three-dimensional finite element analysis. Clin Oral Implants Res ,2004,15 (4) :401-412.
  • 8Hermann JS, Schoolfield JD, Nummikoski PV, et al. Crestal bone changes around titanium implants:a methodologic study comparing linear radiographic with histometric measurements. Int J Oral Maxillofac Implants ,2001,16 (4) :475-485.
  • 9Vela-Nebot X, Rodriguez-Ciurana X, Rodado-Alonso C, et al. Benefits of an implant platform modification technique to reduce crestal bone resorption. Implant Dent,2006,15 ( 3 ) : 313-320.
  • 10Merz BR, Hunenbart S, Belser UC. Mechanics of the implantabutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants, 2000, 15 (4): 519 -526.

二级参考文献11

  • 1张恒,施斌.粘固剂对种植义齿部件稳定性影响的比较研究[J].口腔医学研究,2006,22(4):406-409. 被引量:9
  • 2Tan PL, Dunne JT Jr. An esthetic comparison of a metal ceramic crown and cast metal abutment with an all - ceramic crown and zirconia abutment: a clinical report[J].J Prosthet Dent, 2004, 91:215 -218
  • 3Glauser R, Sailer I, Wohlwend A, et al. Experimental zirconia abutments for implant - supported single - tooth restorations in esthetically demanding regions: 4 - year results of a prospective clinical study[J]. Int J Prosthodont, 2004, 17:285 -290
  • 4Andersson B, Glanser R, Maglione M, et al. Ceramic implant abutments for short - span FPDs: a prospective 5 - year multicenter study [J]. Int J Prosthodont, 2003, 16:640 -646
  • 5Imanishi A, Nakamura T, Ohyama T, et al. 3-D Finite element analysis of all - ceramic posterior crowns[ J ].J Oral Rehabil, 2003, 30:818 -822
  • 6Eskitascioglu G, Usumez A, Sevimay M, et al. The influence of occlusal loading location on stresses transferred to implant - supported prostheses and supporting bone: A three -dimensional finite element study [J]. J Prosthet Dent, 2004, 91:144 - 150
  • 7Yildirim M, Fischer H, Marx R, et al. In vivo fracture resistance of implant - supported all - ceramic restorations [ J ]. J Prosthet Dent, 2003, 90:325 -331
  • 8Att W, Kurun S, Gerds T, et al. Fracture resistance of single - tooth implant - supported all - ceramic restorations after exposure to the artificial mouth[J]. J Oral Rehabil, 2006, 33:380 -386
  • 9Att W, Kurun S, C, erds T, et al. Fracture resistance of single - tooth implant - supported all - ceramic restorations: an in vitro study[J]. J Prosthet Dent, 2006, 95:111 - 116
  • 10施斌,周铸民.单个种植牙的种植体边缘骨的应力分析[J].口腔医学研究,2003,19(1):63-65. 被引量:5

共引文献18

同被引文献91

引证文献12

二级引证文献42

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部