经皮给药电穿孔仪的性能分析 被引量:6

Characteristic analysis of electroporation apparatus for transdermal drug delivery
下载PDF
导出
摘要 本文分析了一种经皮给药电穿孔仪的性能。在该仪器中,电压发生器将12 V的直流电源电压提升到400 V,并对4个蓄能电容2.2μF、22μF、47μF和100μF充电。在脉冲控制器的控制下,蓄能电容通过Ag-AgC l电极向皮肤放电产生指数衰减脉冲电压。记录了4种电容器放电的脉冲电压,结果表明:当截断系数ε≤1/5,可以获得完整的指数衰减脉冲,并且皮肤电阻为3.57±0.66 kΩ。ε受经皮能量ES和脉冲选择的制约,通过设计并行电阻RI和蓄能电容C,可以使蓄能电容释放的能量效率η最高,ε最小。 The characteristic of an electroporation apparatus for transdermal drug delivery is discussed in the paper. The voltage generator generates DC 400 V from DC 12 V power supply using a DC/DC converter in the apparatus, and charges four storage capacitors 2.2 μF, 22 μF, 47 μF and 100 μF. One of the four capacitors discharges to skin through Ag-AgCl electrodes under the control of a pulse controller, and generates a series of exponential decay pulses on the skin. Four wavefonns of the discharge pulses are recorded. The results show that when truncation coefficient is less than 1/5, complete exponential decay pulse can be acquired, and the skin resistance is 3.57± 0.66 kΩ). The truncation coefficient is restricted by both transdermal energy and pulse protocols. If parallel resistance and storage capacitance in the system are selected optimally, the highest power supply energy efficiency and the lowest pulse truncation coefficient can be obtained.
作者 应磊 包家立
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2008年第1期16-20,共5页 Chinese Journal of Scientific Instrument
基金 国家自然科学基金(50147009 50377039)资助项目
关键词 经皮给药 电穿孔 截断系数 经皮能量 脉冲选择 transdermal drug delivery electroporation truncation coefficient transdermal energy pulse protocol
  • 相关文献

参考文献11

  • 1PRAUSNITZ M R, BOSE V G, LANGER R, et al. Electroporation of mammalian skin: A mechanism to enhance transdermal drug delivery[J]. Proc. Natl. Acad. Sci. USA, 1993, 90: 10504-10508.
  • 2SHARMA A, KARA M, SMITH F R, et al. Transdermal drug delivery using electroporation Ⅰ: Factors influencing in vitro delivery of terazosin hydrochloride in hairless rats [J]. J. Pharm Sci., 2000,89(4) : 528-535.
  • 3CHANG S L, HOFMANN G A, ZHANG L, et al. The effect of electroporation on iontophoretic transdermal delivery of calcium regulating hormones [ J ]. J. Control. Rel. , 2000,66: 127-133.
  • 4REGNIER V, TAHIRI A, ANDRE N, et al. Electroporation-mediated delivery of 3 '-protected phosphodiester oligodeoxynucleotides to the skin [ J ]. J. Control. Rel. , 2000, 67: 337-346.
  • 5RAGSDALE C W. Transfection high voltage controller: US,4750100[ P]. 1988-1-7.
  • 6RAGSDALE C W. Transfection high voltage controller: US,642035 [P]. 1997-1-24.
  • 7RAGSDALE C W. Transfection high voltage controller: US,5656926[ P]. 1997-8-22.
  • 8RAGSDALE C W, TAGLIAMONTE J. Electroporation cell with arc prevention/reduction : US, 6258592 [ P ]. 2001-7-10.
  • 9包家立,梁文权,胡巧红,高建青.经皮给药电穿孔仪的研制[J].仪器仪表学报,2000,21(1):66-69. 被引量:10
  • 10VANBEVER R, LECOUTURIER N, PREAT V. Transdermal delivery of metoprolol by electroporation [ J ]. Pharm. Res., 1994,11(11): 1657-1662.

二级参考文献2

共引文献12

同被引文献121

引证文献6

二级引证文献67

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部