X线照片斑点的自相关函数的测试探讨 被引量:3

Research on testing of ACF of radiographic mottling
原文传递
导出
摘要 目的:测试X线照片斑点的自相关函数(auto-corelationfunction,ACF)。材料与方法:对Fuji(CaWO4),Fuji(BaFClEu),天津X线胶片(CaWO4)和天津X线胶片(BaFClEu)4种屏-片组合系统,在同一摄影条件下获得X线照片,用显微密度计对其进行扫描取得X线照片的密度分布曲线D(X),作为离散型随机信号来处理,计算出4种屏-片系统的X线照片斑点的ACF值。结果:与BaFClEu组合的X线照片斑点的ACF值比与CaWO4组合的屏片系统的大。结论:BaFClEu增感屏因结构颗粒大,对X线照片影响大(或相关性强),控制X线照片斑点ACF值的主要措施是,使增感屏的颗粒不宜过大。 Purpose: To test ACF(auto correlation function) of granules on radiograph. Materials and methods: With four kinds of screen film systems Fuji/CaWO 4, Tianjin/CaWO 4, Fuji/(BaFCl∶Eu) and Tianjin/(BaFCl∶Eu), radiographs were obtained under the same exposure conditions. Through sanning these radiographs with microdensitometer the density distribution curves D(x) were obtained. There curves could be recognized as discrete random signals. By processing the curves, the ACF values of the radiographic mottling of these films were calculated. Results: the ACF of BaFCl∶Eu was greater than that of CaWO 4. Conclusion: The larger the granules of film screens, the greater the value of ACF. The method of controlling ACF is to reduce the size of screen granules.
出处 《中华放射学杂志》 CAS CSCD 北大核心 1997年第7期470-473,共4页 Chinese Journal of Radiology
关键词 X线胶片 X线增感屏 质量控制 斑点 医用X线 X ray film X ray intensifying screen Quality control
  • 相关文献

参考文献2

  • 1杨福生,生物医学信号处理,1993年,23页
  • 2杨福生,生物医学信号处理,1993年,25页

同被引文献47

  • 1袁聿德,冯树理,李月卿,朱延好,于兹喜.模糊数学在医学影像技术中的应用简介[J].中华放射学杂志,1995,29(1):11-14. 被引量:5
  • 2袁聿德,冯树理,李月卿,朱延好,孙勇,许慧.X线照片均方根值粒状度的测试探讨[J].中华放射学杂志,1995,29(12):862-865. 被引量:3
  • 3KENNETHR Castleman.Digital Image Processing[M].北京:清华大学出版社,1998..
  • 4BenA. Arnold, PaulO. Scheibe, "noiscanalysisofadigitalradiology system" [J]. Am. J. Roent, 1984, 142, 609-613.
  • 5K Faulkner and B M Moores, "Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function"[J]. Phys. Med. Biol, 1984, 29, 1343-1352.
  • 6Marie Fokey Kijewski and Philip F Judy, "The noise power spectrum of CT images" [J]. Phys. Med. Biol, 1987, 32, 565-575.
  • 7Maryellen Lissak Giger, Kunio Doi, and Charles E. Metz, "Investigation of basic imaging properties in digital radiography[J].2. Noise Wiener spectrum", Med. Phys, 1984, 11,797-805.
  • 8W. Hillen, U. Schiebel, and T. Zaengel, "Imaging performance of a digital storage phosphor system"[J]. Med. Phys, 1987, 14,744-751.
  • 9Ulrich Neitzel, Ingo Maaxk, and Susanne Gunther-kohfahl,"Image quality of a digital chest radiography system based on a selenium detector" [J] .Med. Phys, 1994, 21,509-516.
  • 10James T. Dobbins Ⅲ, "Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systems"[J].Med. Phys. 1995, 22,171-181.

引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部