浮动基准法无创血糖浓度检测技术 被引量:5

Floating-Reference Method of Noninvasive Blood Glucose Concentration Measurement
下载PDF
导出
摘要 介绍了浮动基准法无创伤测量血糖浓度的原理,并模拟研究了其关键因素——径向检测基准位置的存在条件.在径向检测基准位置,漫反射光对葡萄糖浓度变化不敏感,因而可以作为内部基准来消除人体生理背景变化的干扰.通过设计不同背景条件下的Intralipid溶液测量实验,进行了浮动基准法的初步应用实验.实验结果表明,采用浮动基准法进行光谱数据处理后,有效地减小了测量环境变化及Intralipid样品背景变化的影响,极大地增强了葡萄糖浓度的分辨能力. The principle of the noninvasive blood glucose concentration measurement with floating-reference method was introduced. By Monte Carlo simulation, the existing condition of the radial reference location was studied, which was the key factor of the floating-reference method. At the reference location, the diffuse reflectance light is not sensitive to the variation of glucose, thus it can be used as an internal reference to diminish the influence of the physiological background change. Using the Intralipid solutions with different background, the preliminary experimental validation of the floating-reference method was conducted. The results show that after the spectra data are processed by the floating-reference method, the influence induced by the change of measuring circumstance and the variation of Intralipid background is diminished effectively, and the resolving capability on glucose concentration is greatly enhanced.
出处 《纳米技术与精密工程》 EI CAS CSCD 2007年第4期298-301,共4页 Nanotechnology and Precision Engineering
基金 "十一五"国家科技支撑计划重大项目(2006BAI03A03) 中国博士后科学基金资助项目(20060390661)
关键词 浮动基准法 径向基准位置 血糖浓度 floating-reference method radial reference location blood glucose concentration
  • 相关文献

参考文献9

  • 1[1]Chung J,Fan K,Wong T,et al.Method for predicting the blood glucose level of a person.USA,20060253008[P].2006-09-11.
  • 2[2]Khalil O S.Non-invasive glucose measurement technologies:An update from 1999 to the dawn of the new millemium[J].Diabetes Tedmol Ther,2004,6(6):660-697.
  • 3[3]Xu Kexin,Qiu Qinjun,Jiang Jingying,et al.Non-invasive glucose sensing with near-infrared spectroscopy enhanced by optical measurement eonditions reproduction technique[J].Optics and Lasers in Engineering,2005,13(10):1096-1106.
  • 4[4]Mamo K,Tsurugi M,Tamura M,et al.In vivo noninvasive measurement of blood glucose by rlear-hafrared diffuse-reflectance spectroscopy[J].Appl Spearosc,2003,57(10):1236-1244.
  • 5[5]Burmeister J J,Amold M A,Small G W.Noninvasive blood glucose measurements by near-infrared transmission spectroscopy across human tongues[J].Diabetes Technol Ther,2000,2(1):5-16.
  • 6[7]Luo Yunhan,An Lin,Ma Zhen,et al.Discussion on floalingreference method for noninvasive measurement of blood glucose with near-infrared spectroscopy[C]∥Progress in Biomedical Optics and Imaging(Proceedings of SPIE).California USA,2006:60940K.
  • 7[8]Chen Wenliang,Ma Zhen,An Lin,et al.Applying the floatingreference method to improve the precision of noninvasive blood glucose measurement[C]∥Progress in Biomedical Optics and Imaging(Proceedings of SPIE).California,USA,2007:64450M.
  • 8[9]Wang LH,Jacques S L,ZhengL Q.MCML-Monte carlo modeling of light transport in multi-layered lissue[J].Computer Methods and Programs in Biomedicine,1995,47:131-146.
  • 9[10]Matuo K,Tsmugi M,Chin J,et al.Noninvasive blood glucose assay using a newly devdoped near-infrared system[J].IEEE.J Sel Topics in Quan Electronics,2003,9(2):322-330.

同被引文献64

引证文献5

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部