摘要
目的:应用彩色多普勒超声(CDU)预选动脉导管未闭(PDA)栓塞适应证,减少塞子脱落和动脉损伤并发症,提高PDA堵塞成功率。材料和方法:按Porstmann法PDA栓塞适应证标准,先经CDU检测PDA及股动脉内径,选出符合栓塞适应证的单纯PDA32例进行数字化PDA和股动脉造影回顾性验证超声预选准确性。术者在CDU导向下完成顶塞堵闭PDA全过程,以取得最佳嵌顿位置并完全阻断分流。结果:术前CDU预选准确性96.8%,术中CDU引导成功率100%,31例PDA行栓塞术均一次成功,成功率100%,其中21例术后远期CDU随访均未见塞子脱落和再通。结论:术前CDU选择栓塞适应证准确可靠,术中CDU定位法优于X线定位法。
Purpose: Color doppler ultrasound (CDU) was used to select the indication, to diminish the possibility of the plug exfoliation and the arterial injnry and to increase the chance of successful embolization in patients with patent ductus arteriosus (PDA). Materials and Methods:According to standard Porstmann's criterion and 32 cases of simple PDA conforming to embolization indication were selected to undergo digital PDA and femoral arterial angiography for verifing the accuracy of Ultrasonic preconcentration. The complete pocess of prlaging and blocking of PDA was directed by CDU in order to achieve the very best location of wedged plugs and to block the shunts completely. Results: The accuracy of predictive selection of PDA in preoperation examination is 96.8%, the rate of successful guidance by CDU in operation was 100%, 31 cases of embolization of PDA were succeeded in the first time of operation, the rate of success is 100%, 21 cases among them had no sign of postoperation plug exfoliation or recanalization in follow up survey. Canclusion: The selection of the embolization indications by CDU was accurate and reliable. The location of PDA by CDU was superior to x ray methods.
出处
《中国医学影像学杂志》
CSCD
1997年第2期84-86,共3页
Chinese Journal of Medical Imaging
关键词
彩色多普勤超声
动脉导管未闭
栓塞
color Doppler ultrasound
patent ducts arteriosus
embolization