表面三维重建超声心动图测定狗左心腔及塑料容器容量的研究 被引量:2

Echocardiographic Study on Volume Determination of Canine Left Ventricle and Plastic Vessel with Surface Three Dimensional Reconstruction
下载PDF
导出
摘要 目的:观察评价表面三维超声心动图重建系统对容量测定的准确性。材料和方法:采用表面三维重建系统及常用二维超声心动图方法,测定8个塑料容器及10个狗左心室腔容量并与实测值进行比较。结果:各种超声心动图方法所测容量与实测值之间无显著差别,均为高度相关。对所用各种超声方法所测容量值进行方差分析,P>0.05。结论:表面三维超声心动图在容量测定的准确性方面与常用的二维超声心动图方法相近,其优点是操作简便、实用、可形象地显示被测物体的表面轮廓。 Purpose: To evaluate the accuracy of volume measuring with surface three dimensional echocardiography. Materials and Methods: Volume of eight plastic container and ten canine left ventricle were determined by three dimensional echocardiography, a surface wire frame three dimensional technique, and common two dimensional echocardiography methods. The volume obtained from echocardiography was compared with the real volume determination. Results: There were high correlation and no significant difference between each ultrasound method and real measurement. The analysis of variance showed there were not significant difference among each ultrasound method (P>0 05). Conclusion: The accuracy about volume determination got from three dimensional echocardiography and the common two dimensional echocardiography was very close. The merit of three dimensional echocardiography is easy to perform and can get a vivid surface of an object.
出处 《中国医学影像学杂志》 CSCD 1997年第2期110-111,114,共3页 Chinese Journal of Medical Imaging
关键词 三维超声心动图 容量测定 左心室 超声心动图 three dimensional echocardiography volume measurement
  • 相关文献

参考文献2

二级参考文献1

  • 1David J. Skorton M.D.,Steve M. Collins. Quantitation in echocardiography[J] 1987,Cardiovascular and Interventional Radiology(6):316~331

共引文献7

同被引文献29

  • 1张秀芳.生命科学中的物理问题讲座第五讲穆斯堡尔效应及其在生物医学中的应用[J].物理,1994,23(5):307-312. 被引量:2
  • 2李美亚,张之翔.X射线的发现及其对现代科学技术的影响──纪念伦琴发现X射线100周年[J].物理,1995,24(8):474-482. 被引量:10
  • 3Roberts L.3D and stereoscopic observation of bodystructure by ultrasound.J Appl Physiol,1956,9:304.
  • 4Dekker DL,Piziali RL,Deng E Jr.A system for ultrasonically imaging the human heart in three dimensions.Comput Biomed Res,1974,7:544-533.
  • 5Riccabona M,Nelson TR,Pretorid H,et al.Three-dimensional ultrasound: accuracy of distance and volume measurements.Ultrasound Obstet Gynecol,1996,7:429-434.
  • 6Rankin RN,Fenster A,Downey DB,et al.Three-dimensional sonographic reconstruction: techniques and diagnostic applications.Am J Roentgenol,1993,161:695-702.
  • 7Chang FM,Hsu KF,Ko HC,et al.Three-dimensional ultrasound assessments of fetal liver volume in normal pregnancy: a comparison of reproducibility with two-dimensional ultrasound and a search for a volume constant. Ultrasound Med Biol,1997,23:381-389.
  • 8Hughes SW,D′Arcy TJ,Maxwell DJ.Application of a new discrect form of Gauss′theorem for measuring volume.Phys Med Biol,1996,41:1809-1821.
  • 9Liess H,Roth C,Umgelter A,et al.Improvements in volumetric quantification of circumscribed hepatic lesions by three dimensional sonography.Gastroenterol,1994,32:488-492.
  • 10Ivan ES,Chan L,Uerpairojkit B,et al.Growth of fetal forehead and normative dimensions developed by three-dimensional ultrasonographic tecnology.J Ultrasound Med,1997,16:401-406.

引证文献2

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部