期刊文献+

TR-R1根管长度测量仪临床应用的比较研究

The comparative study of TR-R1 endometer in clinical accuracy
下载PDF
导出
摘要 目的对TR-R1根管长度测量仪与RootZX测量仪进行临床应用的比较研究,以评价分析其测量根管长度的准确性。方法将220个患牙的455个根管分别用两种根管测量仪确定根管工作长度,再按测量长度常规充填后进行X线摄片测量,比较二者的准确率。结果TR-R1测量的准确率为95.4%;RootZX测量的准确率为97.1%。结论TR-R1根管长度测量仪的准确率与RootZX一致,应用均安全、可靠,准确率高。 Objective To evaluate and analyze the clinical accuracy of TR-R1 endometer comparing with Root ZX endometer. Method To detect the length of 455 root cannal in 220 cases using TR-R1 and Root ZX respectively and to compare the precise rate with X-ray measuring result. Result The precise rate of TR-R1 is 95.4%, while Root ZX 97.1%. Conclusion Clinical application of TR-R1 in the root cannal working length measurement is reliable and safe with a high accuracy.
作者 贺冰
出处 《福建医药杂志》 CAS 2008年第1期15-17,共3页 Fujian Medical Journal
关键词 根管长度测量仪 TR—R1 ROOT ZX 准确性 Endometer TR-R1 Root ZX Accuracy
  • 相关文献

参考文献5

二级参考文献36

  • 1顾永春,皮昕,周培刚,刘淑香.3803个恒牙侧副管的解剖形态研究[J].临床口腔医学杂志,2004,20(6):334-338. 被引量:16
  • 2[1]Sunada I.J Dent Res,1962,41(5):375-385.
  • 3[2]Huang L.J Endod,1987,13 (2):60-64.
  • 4[3]Lucena-Martin C,Robles-Gijon V,Ferrer-Luque CM,et al.J Endod,2004,30(4):231-233.
  • 5[4]Venturi M,Breschi L.Int Endod J,2005,38 (1):36-45.
  • 6[5]Haffner C,Folwaczny M,Galler K,et al.J Dent,2005,33 (8):619-625.
  • 7[6]Hoer D,Attin T.Int Endod J,2004,37(2):125-131.
  • 8[7]Tselnik M,Baumgartner JC,Marshall JG.J Endod,2005,30 (7):507-509.
  • 9[8]Kaufman AY,Keila S,Yoshpe M.Int Endod J,2002,35 (2):186-192.
  • 10[9]Jenkins JA,Walker WA,Schindler WG,et al.J Endod,2001,27 (3):209-211.