摘要
目的:研究建立一种经皮可释放性弹簧栓栓堵治疗动脉导管未闭(PDA)的新方法。方法:经主动脉造影测定PDA内径及长度,在X线和超声心动图监测引导下,经心导管将可释放性弹簧栓栓堵于PDA。分别采用Jackson弹簧栓(J组)和PDA弹簧栓(P组)。经临床、主动脉造影和超声心动图确定疗效。结果:采用本法治疗21例,PDA平均最狭窄处内径3.18±0.62mm(范围1.5~4.7mm,其中81%超过3mm),平均长度5.17±2.13mm。20例获得成功且无并发症,术后经临床、主动脉造影和超声心动图检查,无残余分流。其中J组8例均获成功;P组13例,成功12例、失败1例。结论:本法属于安全可靠的非开胸治疗PDA的新方法,创伤性很小,值得使用。
Objective:To develop a new clinical method of percutaneous transcatheter closure of patent ductus arteriosus(PDA) using detachable coil. Methods:After the internal diameter and length of PDA were estimated by aortography,a set of detachable coil(Jackson coil for J Group and PDA coil for P Group) was delivered,released and lodged in the PDA through a catheter engaged retrogradely,under the monitoring of echocardiography and X ray. The results were comfirmed by clinical,aortographic and echocardiographic studies. Results:Twenty one patients were eligible.The narrowest internal diameter and length of PDA were 3 18±0 62 mm(range 1 5—4 7 mm,81%>3 mm) and 5 17±2 13 mm,respectively.Among them,20 cases were successfully treated using detached coils without any complications.There were no clinical,angiographic and echocardiographic evidences of residual shunting,including 8/8(100%)in J Group and 12/13(92%)in P Group. One case in P Group was failed. Conclusions:The study shows that the transcatheter coil closure of PDA is a new ,safe and less invasive method for the treatment of patients with PDA,so it is worthy to be used.
出处
《中国循环杂志》
CSCD
北大核心
1997年第5期372-374,共3页
Chinese Circulation Journal
关键词
先天性心脏病
动脉导管未闭
经皮导管栓堵术
Congenital heart disease
Patent ductus arteriosus
Transcatheter closure