期刊文献+

DS5和DCA2000糖化血红蛋白仪检测糖化血红蛋白的比较 被引量:2

下载PDF
导出
摘要 目的:对DS5低压阳离子交换色谱法联合梯度规避法和DCA2000免疫法两种糖化血红蛋白分析仪的不同分析方法进行方法学评价。方法:用低压阳离子交换色谱法联合梯度规避法(X)和免疫法(Y)测定糖化血红蛋白,进行回收率、精密度、干扰因素、相关性及参考范围的分析。结果:两种方法的平均回收率均在95%~104%以内,批内、批间的平均变异系数(CV)均〈5%;10.2mmol/L三酰甘油、8.0mmol/L胆固醇、10.1%HbF对两种仪器均无干扰,684umol/L总胆红素对DS5的层析柱有破坏作用,对DCA2000无干扰;两者结果密切相关,相关系数r=0.9957(P〈0.01),n=50,直线回归方程Y=1.09X+0.1%;DS5测定糖化血红蛋白结果较DCA2000偏低;DS5和DCA2000两者参考范围分别为4.0%-6.0%和4.3%-6.5%。结论:DCA2000测定结果比DS5偏高。
出处 《实用医技杂志》 2008年第8期1021-1022,共2页 Journal of Practical Medical Techniques
  • 相关文献

参考文献1

共引文献1

同被引文献44

  • 1International Expert Committee. International Expert Committee report on the role of the A1C assay in the diagnosis of diabetes [J]. Diabetes Care, 2009,32 : 1327-1334.
  • 2Stratton IM, Alder AI, Nell HA, et al. Association of glycaemic with macrovascular and microvascular complications of type 2 dia- betes ( UKPDS 35 ) : prospective observational study [ J ]. BMJ, 2000,321 (7258) :405-412.
  • 3Rahbar S, Blumenfeld O, Ranney HM. Studies of an unusual he- moglobin in patients with diabetes mellitus[J]. Biochem Biophys Res Commun,1969,36(5) :838-843.
  • 4Kutter D,Thoma J. Hereditary spherocytosis and other hemolytic anomalies distort diabetic control by glycated hemoglobin[J]. Clin Lab, 2006,52 (9-10) : 477-481.
  • 5Asia-Pacific Type 2 Diabetes Policy Group. Type 2 diabetes prac tical targets and Treatment. 4th ed. Sydney: Health Communications Australia Pty Ltd and In Vivo Communications Pty Ltd, 2005.
  • 6Little RR,Rohlfing CL,Wiedmeyer HM, et al. The National Gly- cohemoglobin Standardization Program..A Five-Year Progress Report[J]. Clin Chem,2001,47:1985-1992.
  • 7Bunn FH, Forget BG. Hemoglobin: molecular, genetic and clinical aspects[J]. Philadelphia : WB Saunders Co, 1986 : 425-427.
  • 8Little RR,Roberts WL. A review of variant hemoglobins interfering with hemoglobin Ale measurement[J]. J Diabetes Sci Technol,2009,3(3) :446-451.
  • 9Camargo JL, Felisberto M,Gross JL. Effect of pre analytical vari ables on glycohemoglobin measurements in routine clinical care [J]. Clin Biochem,2004,37(9) :836-839.
  • 10Bry L, Chem PC, Sacks DB. Effects of hemoglobin variants and chemically modified derivatives on assays for glycohemoglobin [J]. Clin Chem, 2001,47 : 153.

引证文献2

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;