期刊文献+

尿液干化学分析的问题探讨 被引量:2

Discussion about the urine dry chemistry analysis
下载PDF
导出
摘要 目的:探讨尿液干化学分析结果正常时与显微镜检查结果之间的关系。方法:采用尿液分析仪检测,再对91例正常标本和98例干化学结果异常的标本进行尿沉渣显微镜检查。结果:尿液分析仪检测结果阴性的标本中,有24例尿沉渣显微镜检查结果阳性;检测结果阳性的标本中,有25例尿沉渣显微镜检查结果阴性。结论:尿液干化学分析不能替代显微镜检查,否则会出现结果假阴性,或假阳性结果而造成试验误差甚至延误疾病的诊断。 Objective to study the relationship between the result of urine dry chemistry analysis and the result of examination under microscope. Method The urine analyzer were applied to examine, and then microscope to examine urine sediment of 91 normal specimens and 98 unusual specimens. Result:Among the negative specimens examined by urine analyzer, 24 specimens were examined positive by microscope; among the positive specimens examined by urine analyzer, 25 specimens were examined negative by microscope. Conclusion: The urine dry chemistry analysis can't take the place of the examination under microscope, or there will be the result that is false negative or false positive ,which can cause the failure of experiment, even delay the diagnosis of disease.
作者 陶良萍 陈玲
出处 《安徽卫生职业技术学院学报》 2008年第2期98-98,91,共2页 Journal of Anhui Health Vocational & Technical College
关键词 尿液干化学分析仪 尿沉渣检测分析仪 假阴性 假阳性 The urine dry chemistry Analyzer, the urine sediment examination Analyzer , False-negative, false positive