数字X线机常规输出稳定性的测试

Stability measurement of digital X-ray unit on their daily output
下载PDF
导出
摘要 目的:通过测试数字成像设备的X线机管电压和曝光时间稳定性,探讨数字成像摄影条件的选择原则。方法:在证明四台数字X线机mAs稳定性合格的前提下,选择合适的kVp档和时间档,测量其高压不准确度和曝光时间不准确度,并同国家标准进行比较。结果:四种X线机管电压不准确度在各个kVp档上均小于5%;时间不准确度大约在20ms档以上小于5%,在20ms档以下大于5%。结论:应用数字X线机摄影时曝光时间不可选择太小。 Objective-To obtain suitable digital radiologic exposure factors of CR and DR by measuring stability of their X-my tube voltage and exposure time. Methods:Four X-ray units were studied, the actual X-ray tube voltage was orderly measured when pre-serdng voltage was 45, 60, 80, 100, 120, the actual exposure time was orderly measured when pre-set exposure time was 2, 6, 40, 50, 100, 500, 1000. By comparing actual measured values and pre-set ones, inaccuracy was obtained. Results:For all of four X-ray units, inaccuracy of voltage were less than 5% ; inaccuracy of exposure time were less than 5% when pre-set time 〉 20ms, and more than 5% when pre-set time 〈 20ms. Conclusion:In practice exposure time should keep high level to ensure X-ray unit output steady.
出处 《医学影像学杂志》 2008年第4期426-428,共3页 Journal of Medical Imaging
关键词 X线机管电压 曝光时间 质量控制 质量保证 X-ray tube voltage exposure time quality control quality assurance
  • 相关文献

参考文献4

二级参考文献29

  • 1吴毅,张建.RMPX射线诊断设备质量控制检测箱的研制[J].中华放射医学与防护杂志,1993,13(5):300-303. 被引量:4
  • 2薛永库,张登森.医用诊断X射线机质量保证中高压,输出量和曝光时间的病?[J].中华放射医学与防护杂志,1993,13(3):147-150. 被引量:1
  • 3范六一,李连波,丁崇海,候金鹏,邓大平.医用诊断X射线质量控制的试点调查[J].中华放射医学与防护杂志,1994,14(3):201-203. 被引量:7
  • 4韩良兵,周淑玲.X线机修理后的质量保证问题[J].中华放射学杂志,1995,29(11):772-772. 被引量:1
  • 5Kenneth A. Fetterly, Nicholas J. Hangiandreou, Beth A.Schueler, and E. Russell Ritenour. Measurement of the presempled two-dimensional modulation transfer function of digital imaging systems[J]. Med. Phys, 2002, 29(5):913-921.
  • 6Minoru Sonoda, Masao Takano, et al. Computed Radiography Utilizing Scanning Laser Stimulated Luminescence[J]. Radiology, 1983, 148: 833-838.
  • 7Maryellen Lissak Giger, and Kunio Doi. Investigation of basic imaging properties in digital radiography. 1. Modulation trausfer function[J]. Med. Phys., 1984, 11(3):287-295.
  • 8Maryellen Lissak Giger, Kunio Doi, and Charles E. Metz. Investigation of basic imaging properties in digital radiography. 2.Noise Wiener spectrum[J]. Med. Phys. 1984, 11(6) :797-805.
  • 9I. A. Cunningham and A. Fenster. A method for modulation transfer function determination from edge profiles with correction for finite-element differentiation[J]. Med. Phys, 1987, 14(4):533-537.
  • 10W. Hillen, U. Schiebel, and T. Zaengel. Imaging performance of a digital storage phosphor system[J]. Med. Phys., 1987, 14(5) :744-751.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部