摘要
本文利用生物胶作为稳定剂和基体改进剂,把难溶环境固体样品制备成悬浊液,直接进样测定煤飞灰中的铅,得到了准确的结果。找出了日立180-80石墨炉原子吸收光谱仪的最佳线性范围7.0~66.7ng/mL,原子化适宜温度2000℃,生物胶最佳用量范围0.2~1.2‰,并分析了酸度、搅拌、粒度对测定结果的影响。
Biochemical was chosen as the stability and improvement reagent to make insoluable solid into slurry,which could be sampled directly to determine Pb in flying coal ash,one of the environmental samples.The optimal conditions of Hitichi GTAAS 180 80 were decided by experiments,such as linear range 7 0—66 7ng/mL atomization temperature 2000℃ and amount of biochemical 0 2—1 2‰.The effects of acid,stirring and particle concentration were also addressed.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
1997年第5期98-102,共5页