XE-2100血液分析仪测定有核红细胞的评估 被引量:3

Performance evaluation of nucleated red blood cell count using the Sysmex XE-2100 automated hematology analyzer
下载PDF
导出
摘要 目的对XE-2100血液分析仪测定有核红细胞的性能进行评估。方法通过测定高、中、低值有核红细胞标本评析其重复性、线性范围、携带污染率和稳定性,并对114份7 d内的新生儿静脉血或脐带血标本作常规测定,评价其对有核红细胞的定性报警、定量测定,并与手工涂片法进行对照。结果XE-2100血液分析仪测定有核红细胞的重复性较好,线性范围较广,携带污染率较小,在3 d内稳定性较好。以手工涂片、染色显微镜油镜复检为金标准,其有核红细胞报警的灵敏度为100%,特异度为76.7%。定量计数时与手工法的相关性良好,R2=0.997 8。结论XE-2100血液分析仪对有核红细胞的测定能应用于临床。 Objective To evaluate the performance of Sysmex XE-2100 automated hematology analyzer to count nucleated red blood cell (NRBC). Methods Precision, linearity, carryover and stability were performed by assaying the high, middle, low NRBC samples. The results of 114 neonatal blood or umbilical cord blood samples measured by the Sysmex XE-2100 were compared with manual. Results The System XE-2100 results showed excellent precision linearity and stability. The specificity for NRBC was 76.7% , and sensitivity was 100%. The correlation was excellent between the System XE-2100 and manual(R^2 = 0. 997 8). Conclusion The performance of NRBC count using the Sysmex XE-2100 automated hematology analyzer is adequate for clinical demand.
出处 《现代医学》 2008年第2期112-114,共3页 Modern Medical Journal
关键词 血液分析仪 有核红细胞 手工分类 报警 hematology analyzer nucleated red blood cell manual differentiation warning
  • 相关文献

参考文献9

  • 1Igout J,Fretigny M,Vasse M,et al.Evaluation of the coulter LH 750 haematology analyzer compared with flow cytometry as the reference method for WBC,platelet and nucleated RBC count[J].Clin Lab Haematol,2004,26(1):1-7.
  • 2束国防,高茂馗,芦慧霞.XE-2100血液分析仪对异常细胞报警的评估[J].东南大学学报(医学版),2005,24(2):98-100. 被引量:10
  • 3Ghosh B,Mittal S,Kumar S,et al.Prediction of perinatal asphyxia with nucleated red blood cells in cord blood of newborns[J].International Journal of Gynecology and Obstetrics,2003,81(3):267-271.
  • 4Perri T,Ferber A,Digli A,et al.Nucleated red blood cells in uncomplicated prolonged pregnancy[J].Obstetrics and Gynecology,2004,104(2):372-376.
  • 5Baschat A A,Gembruch U,Reiss I,et al.Neonatal nucleated red blood cell count and postpartum complications in growth restricted fetuses[J].Journal of Perinatal Medicine,2003,31(4):323-329.
  • 6Stachon A,Kempf R,Holland L T,et al.Daily monitoring of nucleated red blood cells in the blood of surgical intensive care patients[J].Clinica Chimica Acta,2006,366(1-2):329-335.
  • 7Axel S,Nina S,Matthias I,et al.Nucleated red blood cells indicate high risk of in-hospital mortality[J].Journal of Laboratory & Clinical Medicine,2002,140(6):407-412.
  • 8Stachon A,Holland L T,Krieg M.High in-hospital mortality of intensive care patients with nucleated red blood cells in blood[J].Clinical Chemistry and Laboratory Medicine,2004,42(8):933-938.
  • 9Tsuji T,Hamaguchi Y,Wang F S,et al.New rapid flow cytometric method for the enumeration of nucleated red blood cells[J].Cytometry,1999,37(4):291-301.

二级参考文献5

  • 1GRIMALDI E, SCOPACASA F. Evaluation of the Abbott CELLDYN 4000 hematology analyzer[ J].Mn J Clin Pathol, 2000, 113 (4) :497-505.
  • 2PICARD F, GICQUEL C, MARNET L, et al. Preliminary evaluation of the new hematology analyzer COULTER GEN-S in a university. hospital[J]. Clin Chem Lab Med, 1999,37(6):681-686.
  • 3AMOUROUX I, BALAY M, MARFAING K A. Evaluation of the blood analyzer Beckman Coulter LI-I 750:analytic performance, decision rules[ J ]. Ann Biol Clin (Paris), 2003,61 ( 5 ) : 576-584.
  • 4HERKLOTZ R,HUBER A R. Precision and accuracy of the leukocyte differential in the sysmex XE-2100[J]. Sysmex J Int,2001,11(11) :8-21.
  • 5张杰,徐以南,钱敏,陆琳,童明庆.ABX Pentra120血液分析仪性能评估[J].临床检验杂志,2003,21(6):382-383. 被引量:4

共引文献9

同被引文献26

引证文献3

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部